4.1.3 Connecting the Gas Supply
1
Connect the gas supply hose to the gas supply input (T) in Figure 2 of
the e600.
2
Connect the other end of the hose to the pressure outlet of the pressure
regulator or wall outlet of piped medical oxygen system.
3
Turn cylinder valve slowly and fully.
Warning
Extra care must be taken when handling oxygen:
o
The e600 must only be used with medical oxygen.
o
Only use approved medical oxygen compressed gas cylinders.
o
Always begin use with a full oxygen cylinder.
o
Secure oxygen cylinders so they do not fall over.
o
Keep away from excessive heat to avoid the risk of explosion.
o
Do not grease or lubricate oxygen fittings, cylinder valves and
pressure reducers, and do not handle with greasy hands to avoid
the risk of fire.
o
Only open or close cylinder valves by hand or with the correct
cylinder wrench. Open the valve slowly and fully. Do not use any
other tools.
o
Do not smoke or work in areas where open flames are present.
Oxygen supports combustion and exacerbates fires.
o
Only use a pressure reducer with an overpressure relief valve to
limit the delivery pressure in case of a regulator failure!
o
To avoid ventilator malfunction do not attach the ventilator to a
flow control valve or flow meter.
Summary of Contents for e Series
Page 2: ...e600 User Manual 15PL1010 Rev 7 28_8_2017 English version ...
Page 13: ......
Page 14: ......
Page 18: ......
Page 29: ...25 e600 User Manual 15PL1010 Rev 7 28_8_2017 English version Figure 11 SIMV waveform ...
Page 46: ......
Page 48: ......
Page 50: ......
Page 55: ...e600 User Manual 15PL1010 Rev 7 28_8_2017 English version ...