To activate pause function, proceed as follows:
1.
Press Pause/ Resume button (H). The Pause symbol will flash on the
screen with the confirmation symbol
and the Confirmation indicator
(L) to guide users to activate pause function by pressing the Control
Selection Knob (M).
2.
The symbol will be flashing for 10 seconds and then disappear if the
Control Selection Knob (M) is not selected. Users can also press the
Cancel button (K) to quit this selection before 10 seconds.
3.
Once activated, a flashing yellow pause symbol will be displayed on the
screen and ventilator will stop ventilating.
Note
a.
During Pause, there will be an audible alarm associated with the
flashing yellow Warning indicator (A) in Figure 1 every 15 seconds.
Users can press Alarm Silence button to disable the audible alarm
for 2 minutes but the yellow Warning indicator will continue
flashing every 15 seconds.
b.
During Pause, users can change and confirm new ventilation
settings but no ventilation will occur unless Pause function is
disabled.
4.
To cancel Pause function, press Pause/ Resume button (H) again. The
“Resume” symbol will flash on the screen with the confirmation symbol
and the Confirmation indicator (L) to guide users to resume
ventilation by pressing the Control Selection Knob (M).
5.
When ventilation is resumed, the ventilator will recommence ventilation
with current settings shown on the screen unless new set-up selections
were made.
3.2.9 Manual/ Hold
During exhalation phase, if Manual/ Hold button (E) in Figure 1 is pressed, a
mandatory breath will be initiated and either the flow rate or set pressure
control parameter will be delivered as long as the Manual/ Hold button is
pressed or until I-time setting is achieved.
After I-time if the button still pressed, the ventilator will switch to
inspiratory hold function in which the ventilator will cut the flow but will
keep the exhalation port closed in order to block exhaled gas from going to
ambient resulting in the maintenance of lung pressure.
The maximum inspiratory hold time is 6 seconds. After that time, the
ventilator will switch to exhalation phase by opening airway pressure to
ambient.
Summary of Contents for e Series
Page 2: ...e600 User Manual 15PL1010 Rev 7 28_8_2017 English version ...
Page 13: ......
Page 14: ......
Page 18: ......
Page 29: ...25 e600 User Manual 15PL1010 Rev 7 28_8_2017 English version Figure 11 SIMV waveform ...
Page 46: ......
Page 48: ......
Page 50: ......
Page 55: ...e600 User Manual 15PL1010 Rev 7 28_8_2017 English version ...