background image

 

EMC Statement 
This device needs special precautions regarding EMC and need to be installed and 
put into service according to the EMC information provided in the accompanying 
document.  
Portable and mobile RF communications equipment can affect this device. 
Warning: 

 

The use of accessories, transducers and cables other  than those specified with 
the exception of transducers and cables sold by the manufacturer of this device 
as replacement parts for internal components, may result in increased emissions 
or decreased immunity of this model. 

 

This  device  should  not  be  used  adjacent  to  or  stacked  with  other  equipment 

Guidance and Manufacturer’s declaration can be seen in the attachment.

 

EMC Declaration 

Guidance and Manufacturer’s Declaration 

- Electromagnetic Immunity this device 

is  intended  for  use  in  the  electromagnetic  environment  specified  below.  The 
customer  or  the  user  of  this  device  should  ensure  that  it  is  used  in  such  an 
environment.

 

Immunity Test 

Compliance 

Electromagnetic 
Environment - Guidance 

RF emissions 
CISPR11 

Group 1 

This  device  used  RF  energy 
only  for  its  internal  function.  
Therefore,  its  RF  emissions 
are  very  low  and  are  not 
likely 

to 

cause 

any 

interference 

in 

nearby 

electronic equipment. 

RF emissions 
CISPR11 

Class B 

This device is suitable for use 
in  all  establishments  other 
than  domestic  and  those 
directly  connected  to  the 
public  low  voltage  power 
supply  network  that  supplies 
buildings  used  for  domestic 
purposes. 

Harmonic emmissions 
IEC 61000-3-2 

Not applicable 

Voltage fluctuations/ 
flicker emissions  
IEC 61000-3-3 

Not applicable 

 

 

 

Summary of Contents for TTM-6/0740

Page 1: ...TTM 6 0740 Low Frequency Therapy Instrument Instruction Manual Please read all instructions carefully and retain for future use ...

Page 2: ...portant Safety Information 4 5 Battery Warning Preparation 6 Package Content Parts About Therapy Machine Set up your Device 7 Getting Start After Used Instruction of Use 8 Care and Maintenance 9 Troubleshooting EMC Declaration 10 12 Specification 13 Warranty Information 14 ...

Page 3: ...pretation of measurements or if you have or suspect you have a medical issue Do not use this product on pregnant women children or people without ability to express their own consciousness Do not use it on patients with cardiac demand pacemaker defibrillator or implanted metallic or electronic device Precautions 1 Do not move the instrument while using or when the user wishes to change treated are...

Page 4: ...ople with abnormally sensitive skin heart disease abnormal blood pressure malignant tumor serious cerebrovascular disease acute disease epilepsy or currently under doctor s treatment should consult a doctor before use this product 5 Do not use this product in places where there are high heat inflammables and electromagnetic radiation 6 Do not use this product while bathing or sleeping 7 Do not use...

Page 5: ...t may cause cardiac arrhythmias 19 Stimulation should not be applied over swollen infected or inflamed areas or skin eruptions e g phlebitis thrombophlebitis varicose veins etc Technical description is contained in the instruction manual BATTERY WARNING Do not mix alkaline stand carbon zinc and rechargeable batteries nickel hydride Do not mix old and new batteries Non rechargeable batteries are no...

Page 6: ...t mode and other parameters such as intensity through function buttons shown below on the host and know about the current status through LCD at any time Treatment Mode 5 Full automatic treatment mode and 5 manual treatment mode which satisfy to different demands Electrode Pad To be attached to bare skin to perform stimulation Connecting Cable To be used to make the connection between the machine a...

Page 7: ...g Start Tear off the protective films on electrode pads and then paste the electrode pads onto painful areas Relax your body completely Push the power switch on the right of the host to on position and press On button to power on and start treatment It s recommended to use the instrument twice a day 15 minutes each time Caution Only use electrode pads equipped with the product Do not dispose the e...

Page 8: ...cles of Mode 1 Mode 2 Mode 5 iv Cycles of Mode 2 Mode 1 Mode 4 and v Cycles of Mode 5 Mode 1 Mode 4 that each Mode will work for 30 seconds within each cycle Default setting as Mode 1 After each conversion of treatment mode the intensity will be back to 1st level 5 Press Manual button to choose from 5 manual treatment modes which included Mode 1 to Mode 5 After each conversion of treatment mode th...

Page 9: ...ign object inside the compartment Solution Check remove batteries re install batteries according to polarity LCD Display well but no stimulation Reason Two core connecting cable not connected well OR is there any protective film left on the pad Solution Detach two core connecting cable again OR remove protective film Solution Change the location of electrode pads for Mild allergy user or shorten t...

Page 10: ...chment EMC Declaration Guidance and Manufacturer s Declaration Electromagnetic Immunity this device is intended for use in the electromagnetic environment specified below The customer or the user of this device should ensure that it is used in such an environment Immunity Test Compliance Electromagnetic Environment Guidance RF emissions CISPR11 Group 1 This device used RF energy only for its inter...

Page 11: ...uld be that of a typical commercial or hospital environment 1 kV for input output lines Surge 1 kV line to line Not applicable IEC 61000 4 5 2 kV line to earth Voltage dips short interruptions and voltage variations on power supply input lines IEC610004 11 5 UT 95 dip in UT for 0 5 cycle Not applicable 40 UT 60 dip in UT for 5 cycle 70 UT 30 dip in UT for 25 cyce 5 UT 95 dip in UT for 5 sec Power ...

Page 12: ... distance in meters m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey ᵃ should be less than the compliance level in each frequency range ᵇ Interference may occur In the vicinity of equipment marked with the following symbol Note 1 At 80 MHz and 800 MHz the higher frequency range applies Note 2 These guidelines may not apply in all situations Electromagnet...

Page 13: ... 00056W cm2 500 Ω Pulse Duration 20μs 220μs Frequency 1 Hz 330Hz Default Treatment Time 15min Environment Operations Temperature About 5 Cto 40 C Humidity About 15 to 93 RH Storage Temperature About 25 C to 70 C Humidity About 0 to 93 RH Transport Temperature About 10 C to 40 C Humidity About 15 to 93 RH Barometric Pressure for above environment About 700hPa to 1060hPa To improve performance these...

Page 14: ...No charge will be applicable for such repair or replacement SERVICE AND REPAIR If service is required for this product you should first contact Nuvomed Inc Customer Service at info nuvomed us or by calling Toll Free Number 1 866 815 4714 Monday to Friday 10am to 6pm EST NOTE Manufacturer cannot assume responsibility for loss or damage during incoming shipment As a precautionary measure carefully p...

Reviews: