Roller Base
A 08 149
Gas Changeover Unit for two CO
2
gas tanks (Tanks not included)
F 06 003
0,22
µm biological filter (Spare)
R 11 007
Single Stage Regulator for CO
2
gas tank
R 01 154
Stainless steel shelf
A 08 142 and K 13 009 cannot be used at the same time.
3.
PRECAUTIONS AND LIMITATIONS ON USE
The user shall pay attention to the following:
Do not operate the instrument for purposes other than its main purpose.
The instrument should only be used by authorized and trained staff after the
instruction manual has been read carefully. Only authorized technical staff can
handle the product in case of a failure.
Electricity line should be suitable for the power of device and correctly grounded
power supply should be used.
Only original spare parts and original accessories supplied by Nüve should be used.
The set temperature should not destroy the structure of the samples without user’s
notice.
The boiling points of the samples should be higher than the set temperature.
The freezing points of the samples are lower than the set the temperature.
The samples which may liquefy and expand should not be in a sealed container.
Liquids which may expand during heating should not overflow from their containers.
The safety thermostat should be adjusted to the temperature higher than the set
temperature.
There should not be any material in the chamber that can damage to the device.
The materials which will be subject to heating should not be combustible, explosive,
heat susceptible, flammable, adhesive and fusible.
Ensure that the vapors and gases generated during the operation are not harmful to
human health and flammable or explosive.
CO
2
is a toxic gas. If the door of EC 160 is kept opened for long periods, the
CO
2
gas in the chamber is diffused to the laboratory. In that case, ventilate the
laboratory.