nubeca RT-1261B Owner'S Manual Download Page 1

311-1261000-010   Version 1.0 April, 2009

          RT-1261B Ear Thermometer

Owner’s Manual

Summary of Contents for RT-1261B

Page 1: ...311 1261000 010 Version 1 0 April 2009 RT 1261B Ear Thermometer Owner s Manual ...

Page 2: ...en the thermometer is used by on or near children handicapped persons or invalids 2 Use the thermometer only for the intended use described in this manual 3 Do not use the thermometer if it is not working properly or if it has suffered any damage 4 Do not use accessories which are not supplied or recommended by the manufacturer KEEP THESE INSTRUIONS ...

Page 3: ...TABLE OF CONTENT 06 Cautions and Warnings 07 Restrictions of Use 08 Introduction 10 Package Contents 11 Thermometer Parts 12 LCD Screen 14 Replacing the Battery 16 About Normal Body Temperature Fever ...

Page 4: ...Filter 18 Taking Ear Thermometer 20 Hints on Taking Ear Temperature 21 Recalling the Memory 23 Viewing Results on a Personal Computer 26 Care Cleaning 26 Fever Indicator 27 Troubleshooting 28 Specifications 29 Symbol Information 30 Warranty ...

Page 5: ...gly hitting it with a hard object Only use the lens filter for infrared nu beca RT 1261 ear thermometer For proper hygiene do not share lens filter Damaged lens filter may result in error display Do not disassemble the thermometer Basic safety precautions should always be observed especially when the thermometer is used on or near children and disabled persons This thermometer is not intended to be a...

Page 6: ...e accuracy cannot be ensured when blood or drainage is found in the ear canal Take temperature from the other ear if ear drops or medications have been placed in an ear For a person who wears ear plug or hearing aid remove the device and wait for 15 minutes before taking temperature NOTE Never try to clean inside the ears You may accidentally damage the eardrum or its surrounding tissues Remove ex...

Page 7: ... to measure temperature instantly nu beca RT 1261 ear thermometer is an elegantly designed infrared thermometer meant for your family Intended use nu beca RT 1261 ear thermometer is intended for the intermittent measurement and monitoring of human body temperature from ear canal The device is intended for use by people of all ages in the home ...

Page 8: ...he thermometer measures the infrared heat generated by the eardrum and its surrounding tissue The thermometer then converts it into a temperature value shown on LCD NOTE The thermometer does not emit any infrared signal ...

Page 9: ...Package Contents Lens Filter Ear Thermometer Two 1 5V AAA Batteries User s Manual Warranty Card nu beca RT 1261 Storage Case ...

Page 10: ...Thermometer Parts 1 1 2 3 4 6 7 8 9 10 5 Lens filter Lens filter ejection button On Memory button Sticker Battery cover Data port Probe lens Probe Lens filter detector Scan button 2 3 4 6 7 8 5 2 9 10 ...

Page 11: ...LCD Screen 1 9 7 2 10 6 3 5 4 8 ...

Page 12: ...r temperature indicator Non lens filter warning Temperature scanning in progress Temperature display Memory mode Record numbers Low bettery indicator Temperature unit Temperature indicator for memory mode Communication symbol ...

Page 13: ...the new battery in the battery compartment and press it in until the battery is firmly secured Replacing the Battery The thermometer comes with two 1 5V AAA alkaline batteries Replace it when appears Please follow the steps to replace new batteries ...

Page 14: ...meter works when appears we still recommend that you change the battery to obtain an accurate result Remove the battery if stored for a long period of time The battery should be kept out of reach of children If they are swallowed promptly see a doctor for help ...

Page 15: ...n overdressing or immunization Some people may not experience fever even when they are ill These include but are not limited to infants younger than 3 months old persons with compromised immune systems persons taking antibiotics steroids or antipyretics aspirin ibuprofen acetaminophen or persons with certain chronic illnesses Please consult your physician when you feel ill even if you do not have ...

Page 16: ...art to test your ear temperature NOTE If a lens filter is not attached on the probe LCD will show until a new lens filter is firmly attached to the probe The used lens filter should be removed after each measurement by a clean and new lens filter to ensure accuracy reading For hygiene use attach a new lens filter each time and do not touch its tip ...

Page 17: ...2 Press and release the On Memory button to turn on the thermometer When ready the thermometer displays the last measurement Step 1 Make sure lens filter is firmly attached to the probe Step 3 Gently fit the probe into the ear canal ...

Page 18: ...s and release the Scan button Do not remove the thermometer until it beeps Step 6 Discard used lens filter into trash by pressing lens filter ejection button NOTE Turn off the thermometer by pressing On Memory button twice It will automatically turn off if left idle for 3 minutes ...

Page 19: ...younger than 3 years old and who have a compromised immune system and the presence absence of fever is critical When you are learning to use the thermometer 2 Do not take a reading while eating and or talking Wait 30 minutes after any of the following situations before taking a measurement When you have your ear covered After exercising swimming or taking a bath When expose to extreme temperature ...

Page 20: ...ometer Step 3 Press the On Memory button for 3 seconds to enter memory mode Each time you press the On Memory button a result will be displayed in the order of dates latest result shown first together with and number from 1 to 10 When the memory is full the oldest result is deleted as the new one added When the last record is displayed in the LCD press On Memory button again to return the first reco...

Page 21: ...Step 4 Exit the memory Press the Scan button and LCD will show latest results with flashing Then press On Memory button again and LCD will show OFF to exit memory ...

Page 22: ...amped with 70 alcohol The body of the thermometer is not water resistant Never put the thermometer under a running tap or submerge it into water Use a soft and dry cloth to clean it Do not use abrasive cleaners Store the thermometer in a cool and dry location Free from dust and away from direct sunlight Fever Indicator When the measurement is equal or over 38 0 red backlight is shown together with...

Page 23: ...hen the batteries can t provide enough power for a test Put the thermometer under operating temperature range of 10 to 40 Put the thermometer under operating temperature range of 10 to 40 Please place probe cover into probe again Review the instructions and re start the measurement procedure If the above steps do not work Please contact the dealer Please replace batteries as soon as possible Pleas...

Page 24: ... 0 96 8 to 102 2 0 3 0 5 from 34 0 to 35 9 93 2 to 96 6 and from 39 1 to 42 2 102 4 to 108 0 10 to 40 50 to 104 95 RH or less 20 to 60 4 to 140 95 RH or less 10 measurements Dimensions Model no RT 1261B Weight Battery External output Display resolution Accuracy Temperature unit Operating temperature range Operating humidity Storage temperature range Storage humidity Memory capacity Displayed tempe...

Page 25: ...or use Keep away from sunlight Keep dry Temperature limitation Date of manufacture Manufacturer Serial number Caution consult accompanying documents Authorized representative in the CE mark Type BF equipment European Community Do not use if package is damaged ...

Page 26: ...the Warranty Certificate and Warranty Registration Card are duly complete with date of purchase serial number and dealer s stamp and if the Warranty Registration Card is sent to local distributor office not later than 14 days from the date of purchase This warranty is void if this product has been repaired or serviced by unauthorized person This warranty does not cover defects caused by misuse abuse...