Fig. 16
Gasket
Non-Retraction Valve
17
FR
EN
NSK Coupling
1) Remove the Coupling Joint from the hose.
2) Remove the back end Gasket.
3) Pull and remove the water tube, and replace the Non-Retraction Valve.
4) Insert the new Non-Retraction Valve securely and remount the back-end Gasket.
*Refer to Spare Parts List to identify the correct parts.