NSK S-Max Series Operation Manual Download Page 3

2

ENGLISH  

GERMAN  FRENCH  SP

ANISH  IT

ALIAN

3

· Check for vibration, noise and overheating outside the patient's oral cavity before use. If any abnormalities are found, stop 

using the handpiece immediately and contact your authorized dealer.

· Galling of bearing and gear or entering abnormal materials into head may cause a burn injury.
· Do not allow any impact on to the handpiece. Do not drop the handpiece.
· Do not attempt to disassemble the handpiece or tamper with the mechanism.
· Make sure you have an additional handpiece ready in case of unexpected malfunction during operation.
· Sterilization in an autoclave is recommended before initial use and thereafter prior to every use.
· Do not connector disconnect the  bur/drill  or handpiece until the motor or has completely stopped.
· Do not use recommended burIf you use that bur/drill, it may bent or fly away during rotation and do not exceed the drive 

motor speed recommended the bur manufacturers. 

· Do not use bent, damaged or sub-standard  burs/drills. And do not exceed the drive motor speed recommended the  bur/drill  

manufacturers. This may cause shank bend or brake even within the recommended speed.

· Always keep the  bur/drill shank clean. Debris in the chuck mechanism can prevent burs from seating properly and cause 

them to fly out during use.

CAUTION

· Do not operate the handpiece without the bur/drill.
· Do not push the push-bottom during operation. If it pressed, it may

-OVERHEATING of the handpiece head. Special caution must be exercised during use to keep cheek tissue AWAY from 

the push-button of the handpiece. Contact with cheek tissue may cause the push-button to depress and injury to the 
patient may occur.

-Result of malfunction.

· Always operate with water spray. Failure to do so may result in overheating.
· During cut the teeth, wear eye protection and dust protective mask for your health.
· Do not use excess pressure on the treatment area.
· After each use, be sure to clean, lubricate and sterilize the handpiece. If blood is left on the handpiece, it can clot causing rust. 

Neglect lubrication may cause OVERHEATING. 

· Do not wipe nor immerse the handpiece or tips in acidic water or acidic solutions.
· Perform periodical maintenance checks.
· If the handpiece has not been used for a long period, check for noise, vibration, and overheating before use.

001

Summary of Contents for S-Max Series

Page 1: ...OM C0425E Surgical Handpiece 09 03 01 M 001 001 OM C0425E S Max SG20 Ti Max Ti Max Ti Max Ti Max X SG20L X SG25L X SG93L X SG93...

Page 2: ...G20 This handpiece is used only for the Surgical Unit with Torque Calibration System such as NSK Surgic XT Plus Using this product with other surgical units may break the handpiece or bone CAUTION Onl...

Page 3: ...the bur drill manufacturers This may cause shank bend or brake even within the recommended speed Always keep the bur drill shank clean Debris in the chuck mechanism can prevent burs from seating prope...

Page 4: ...long surgical burs Do not use burs longer than 26mm The shanks of longer burs may warp under Always use irrigation water during the operation Head may overheat because of the high speed rotation Even...

Page 5: ...nting and Removing Bur Drill 1 Mounting X SG20L X SG20 X SG25L 1 Insert the bur drill into the head 2 Lightly turn the bur drill until it engages with the latch mechanism Fig 2 3 Push bur drill inward...

Page 6: ...ool and operation procedure For X SG25L X SG93L X SG93 it can connect the Irrigation Tube to the External Spray Nozzle firmly 1 External Spray Nozzle Connect the Irrigation Tube to the External Spray...

Page 7: ...r Only a drill with internal irrigation is used 1 Connect the ends of the Y Connector to the External Spray Nozzle and Internal Spray Nozzles refer as detailed in procedures 1 and 2 2 Connect the Irri...

Page 8: ...of the handpiece hold firmly and spray for 2 3 seconds Check the tissue if it is clean the handpiece can be autoclaved if not spray again until clean For optimum lubrication use NSK Care 3 Plus E Type...

Page 9: ...sert into an autoclave pouch Seal the pouch Only if using a vacuum autoclave 4 Autoclavable up to max 135 ex Autoclave for 20 min at 121 C or 15 min at 132 C 5 Keep the handpiece in the autoclave pouc...

Page 10: ...od Optic X SG20L X SG25L X SG93 Clean the Glass Rod Optic with cotton swab soaked in alcohol Fig 12 CAUTION Do not use a pointed tool or sharp edged to clean the Glass Rod Optic They may damage the mi...