NSK S-Max Series Operation Manual Download Page 1

OM-C0425E

Surgical Handpiece

’09.03.01 M

001

001

OM-C0425E

S-Max SG20

Ti-Max

Ti-Max

Ti-Max

Ti-Max

X-SG20L

X-SG25L

X-SG93L

X - S G 9 3

Summary of Contents for S-Max Series

Page 1: ...OM C0425E Surgical Handpiece 09 03 01 M 001 001 OM C0425E S Max SG20 Ti Max Ti Max Ti Max Ti Max X SG20L X SG25L X SG93L X SG93...

Page 2: ...G20 This handpiece is used only for the Surgical Unit with Torque Calibration System such as NSK Surgic XT Plus Using this product with other surgical units may break the handpiece or bone CAUTION Onl...

Page 3: ...the bur drill manufacturers This may cause shank bend or brake even within the recommended speed Always keep the bur drill shank clean Debris in the chuck mechanism can prevent burs from seating prope...

Page 4: ...long surgical burs Do not use burs longer than 26mm The shanks of longer burs may warp under Always use irrigation water during the operation Head may overheat because of the high speed rotation Even...

Page 5: ...nting and Removing Bur Drill 1 Mounting X SG20L X SG20 X SG25L 1 Insert the bur drill into the head 2 Lightly turn the bur drill until it engages with the latch mechanism Fig 2 3 Push bur drill inward...

Page 6: ...ool and operation procedure For X SG25L X SG93L X SG93 it can connect the Irrigation Tube to the External Spray Nozzle firmly 1 External Spray Nozzle Connect the Irrigation Tube to the External Spray...

Page 7: ...r Only a drill with internal irrigation is used 1 Connect the ends of the Y Connector to the External Spray Nozzle and Internal Spray Nozzles refer as detailed in procedures 1 and 2 2 Connect the Irri...

Page 8: ...of the handpiece hold firmly and spray for 2 3 seconds Check the tissue if it is clean the handpiece can be autoclaved if not spray again until clean For optimum lubrication use NSK Care 3 Plus E Type...

Page 9: ...sert into an autoclave pouch Seal the pouch Only if using a vacuum autoclave 4 Autoclavable up to max 135 ex Autoclave for 20 min at 121 C or 15 min at 132 C 5 Keep the handpiece in the autoclave pouc...

Page 10: ...od Optic X SG20L X SG25L X SG93 Clean the Glass Rod Optic with cotton swab soaked in alcohol Fig 12 CAUTION Do not use a pointed tool or sharp edged to clean the Glass Rod Optic They may damage the mi...

Reviews: