36
Operation Procedure
36
<Default Factory Setting>
Mode
Memory No.
Gear ratio
Rotation speed (min
-1
)
Torque (N
・
cm)
Auto reverse mode
General
Application
Mode
1
1:5
200,000
2
1:1
40,000
3
1:1
30,000
4
1:1
20,000
5
1:1
10,000
6
4:1
3,000
Rotary Endo
Mode
1
1:1
250
1.0
AUTO REVERSE FORWARD
2
1:1
250
1.5
AUTO REVERSE FORWARD
3
1:1
250
2.0
AUTO REVERSE FORWARD
4
1:1
250
3.0
AUTO REVERSE FORWARD
5
1:1
300
2.0
AUTO REVERSE FORWARD
6
1:1
1,000
3.0
AUTO STOP
NOTICE
If necessary, take a memo of the latest settings before initializing.
4-8
Overheat Prevention
Following functions detect a risk of overheating from motor current and rotation speed and then slows
down and stops motor rotation by activating the protection circuit to prevent incidents or failure due to the
overheat of the handpiece or the motor.
<Motor Overheat Prevention Function>
This function stops the rotation of the motor when detecting
overload or a risk of abnormal heating of the motor.
”Warning” and ”HE0” will be displayed along with alarm
sound and rotation of the motor will be slowed down.
Continued use will result in motor stop and Error Code “E8” will be displayed. (Refer to “7-1 Error Code”)