30
Operation Procedure
4
After selecting the correct gear ratio, press the
Check Key again to start the checking process.
4
5
The motor automatically starts to operate at low speed and then accelerates to high speed to automatically
measure the state of the handpiece.
CAUTION
Ř
Start the checking process while holding the handpiece. Checking, while leaving the handpiece on the handpiece
stand, may cause drop and damage of the handpiece and injury to user or patient.
Ř
To avoid any problems, care should be taken as the motor operates and rotates automatically.
Ř
During the checking process, the system is designed to stop in an emergency by pressing any key on the control
panel. So, if you sense any danger, press any key on the control panel for an emergency stop.
NOTICE
Checking is possible with or without a bur/file.
Refrain from using burs which are not specified in the instruction of handpiece, otherwise it will lead to wrong
results.
6
Once the process is completed, "Fin" will be
displayed on the screen after a beep sound along
with the <Checking Result> displayed on the
screen as one of these cases: