26
Alarm code
Cause
Solution
AL 12
Temperature out the normal range
Perform the automatic cleaning cycle (§ 6.1).
AL 13
Steam temperature sensor faulty
Call for a technical service
AL 15
Chamber temperature sensor faulty
Call for a technical service
AL 16
Pressure sensor faulty
Call for a technical service
8.5
CLASS B ADDITIONAL ALARMS
TABLE D
Alarm code
Phase involved
Cause
Solution
18
Drying phase
Drying interrupted
Dry the load
31
Drying phase
Vacuum not sufficient
Excess of load