
21
6. MAINTENANCE
6.1
AUTOMATIC PERIODIC CLEANING CYCLE
For a proper operation of your sterilizer it is indispensable that a correct and regular maintenance is carried
out. To this end, it is important to perform a cleaning procedure, as described on the following, at least once
every 15 days or, in case of intense usage, every two water full of the main tank.
For further safety, after 60 cycles carried out without an intermediate cleaning cycle, the equipment
will display the message <NEED CLEANING
>.
IMPORTANT
Take the basket and the trays out of the chamber
, wash with an ordinary dish washing
powder, rinse with water and wipe.
DO NOT USE ABRASIVE PRODUCTS.
WARNING
DO NOT PERFORM THE CLEANING CYCLE WITH TRAYS IN THE CHAMBER
.
Carry out the chamber surface cleaning when the equipment is cold.
NOTE
To open the door and carry out these preliminary operations, it is necessary to switch on
the equipment; ended the maintenance, switch off the unit to avoid excessive heating. To
close the door switch on the unit again.
1.
Put a cleaning tab in the chamber and close the door.
THE CHAMBER MUST BE EMPTY !
2.
Select the cycle
C CLEANING
3.
Push
START
; this cycle takes approx. 15 minutes.
4.
At the end of the cycle,
Ready
will turn on; open the door and wipe
the chamber with a clean cloth slightly soaked with demineralized wa-
ter and pure alcohol.
Do not use sponges, brushes, abrasive steel
wool or paper
.
Thanks to the electronic control system, the number of maintenance cy-
cles actually performed is continuously recorded and updated.
The missing of appropriate and regular maintenance according to the
above guidelines could cause an early and more frequent service activities
and the
lapse of the warranty
.
6.2
CLEANING THE INSTRUMENTS BEFORE THE
STERILIZATION
In order to extend the sterilizer life, we recommend to carry out an accurate cleaning of instruments; one of
the main causes of an early equipment wear is the settlement and accumulation of debris and fragments for
inadequately instrument cleaning, and consequent stains, fouling and progressive clogging of filters, electro-
valves and tubing.