27
Post-use Maintenance
English
[[For handpiece stand]]
[Cleaning]
1
Wipe the surface of the handpiece stand with
a cloth moistened with water�
2
Clean the surface of the handpiece stand
under running water with a soft bristled
brush for more than 20 seconds�
Then rinse the handpiece stand for more
than 5 seconds�
Water temperature: ≤38°C
Water quality:
Water available as
drinking water
Water amount:
≥3�5L/min
3
Wipe the surface of the handpiece stand with
a dry cloth�
4
After cleaning, under appropriate lighting
(≥500lx), inspect the handpiece stand for
contamination�
If any visible contamination remains, repeat
the process from procedure No� 2 until it is
visually clean�
≥
500lx
[Disinfection]
1
Wipe the surface of the handpiece stand with
disinfectant wipes (Minuten Wipes:ALPRO)�
[[For control unit]]
[Cleaning]
1
Remove the barrier film from the control unit�
2
Wipe the surface of the control unit with a
cloth moistened with water�