26
Post-use Maintenance
CAUTION
●
Immediately after cleaning, disinfecting and drying cycle is complete, the contra angle head must
be removed from the washer-disinfector (within 1 hour) to prevent corrosion�
●
After using the washer-disinfector, dry the product completely� Remaining moisture causes internal
corrosion, etc�
●
After cleaning and disinfecting the contra angle head, be sure to perform “6 – 3 Lubrication” and
“6 – 4 Sterilization”�
[[For motor handpiece]]
CAUTION
●
The motor handpiece cannot be cleaned with washer-disinfector�
[Cleaning]
1
Wipe the debris on the surface of the motor
handpiece with a cloth moistened with water�
2
Under appropriate lighting (≥500lx), inspect
the motor handpiece for contamination�
If any visible contamination remains, repeat
the process from procedure No� 1 until it is
visually clean�
≥
500lx
[Disinfection]
1
Wipe the surface of the motor handpiece
with disinfectant wipes (Minuten
Wipes:ALPRO)�
Summary of Contents for ENDO-MATE DT2
Page 1: ......
Page 9: ...8 Product Description 3 3 Control panel...
Page 53: ...52 Produktbeschreibung 3 3 Bedienfeld...
Page 98: ...98 Description du produit 3 3 Panneau de contr le...
Page 142: ...142 Descripci n del producto 3 3 Panel de control...
Page 186: ...186 Descrizione del prodotto 3 3 Pannello di controllo...
Page 230: ...230 Descri o do produto 3 3 Painel de controlo...
Page 267: ......