18
5
Operation Procedure
5 – 1
Standard operation
1
Select a program number� (1-9)
2
Select the rotation direction�
Forward rotation
Reverse rotation
3
Select the gear ratio of the handpiece to be
used�
4
Set the rotation speed recommended by file
manufacturers�
5
Set the torque limit value recommended by
file manufacturers� (Only for the forward
rotation)
2
3
4
5
6
1
(Setting range)
Gear ratio
Rotation speed (min
-1
)
Torque limit (N·cm)
20:1 Reduction
100-650
0�8-6�5
16:1 Reduction
120-800
0�6-5�2
10:1 Reduction
200-1,300
0�4-3�2
4:1 Reduction
500-3,200
0�1-1�3
1:1 Drive
2,000-13,000
- -