Guidance and manufacturer's declaration - electromagnetic immunity

The ENDO-MATE DT is intended for use in the electromagnetic environment specified below. The customer or the user of the ENDO-MATE DT should assure that it is 
used in such an environment.

NOTE:

 

Ut is the a.c. mains voltage prior to application of the test level.

Immunity test

IEC60601 test level

Electromagnetic environment - guidance

Electrostatic discharge (ESD)
IEC61000-4-2

Electrical fast transient/burst
IEC61000-4-4

Surge
IEC61000-4-5

Voltage dips, short 
interruptions and voltage 
variations on power supply 
input lines
IEC61000-4-11

Power frequency (50/60Hz)
magnetic field
IEC61000-4-8

±(2,4) 6 kV contact
±(2,4) 8 kV air

±2kV for power supply lines
±1kV for input/output

±1kV line(s) to line(s)
±2kV line(s) to earth

<5% Ut (>95% dip in Ut)
for 0.5 cycle

40% Ut (60% dip in Ut)
for 5 cycles

70% Ut (30% dip in Ut)
for 25 cycles

<5% Ut (>95% dip in Ut)
for 5 sec

3 A/m

Compliance level

±(2,4) 6 kV contact
±(2,4) 8 kV air

±2kV for power supply lines
±1kV for input/output

±1kV line(s) to line(s)
±2kV line(s) to earth

<5% Ut (>95% dip in Ut)
for 0.5 cycle

40% Ut (60% dip in Ut)
for 5 cycles

70% Ut (30% dip in Ut)
for 25 cycles

<5% Ut (>95% dip in Ut)
for 5 sec

3 A/m

Floors should be wood, concrete or ceramic tile. If floors are 
covered with synthetic material, the relative humidity should 
be at least 30%.

Mains power quality should be that of a typical commercial 
or hospital environment.

Mains power quality should be that of a typical commercial 
or hospital environment.

Mains power quality should be that of a typical commercial 
or hospital environment. If the user of the ENDO-MATE DT 
requires continued operation during power mains 
interruptions, it is recommended that the ENDO-MATE DT be 
powered from an uninterruptible power supply or a battery.

Power frequency magnetic fields should be at levels 
characteristic of a typical location in a typical commercial or 
hospital environment.

Guidance and manufacturer's declaration - electromagnetic immunity

The ENDO-MATE DT is intended for use in the electromagnetic environment specified below. The customer or the user of the ENDO-MATE DT should assure that it is 
used in such an environment.

NOTE 1

 

 

At 80MHz and 800MHz, the higher frequency range applies.

NOTE 2

 

 

 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.

a  Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobiles radios, amateur radio, AM and FM radio 

broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an 
electromagnetic site survey should be considered. If the measured field strength in the location in which the ENDO-MATE DT is used exceeds the applicable RF 
compliance level above, the ENDO-MATE DT should be observed to verity normal operation. If abnormal performance is observed, additional measures may be 
necessary, such as reorienting or relocating the ENDO-MATE DT.

b  Over the frequency range 150kHz to 80MHz, field strengths should be less than 3 V/m.

Immunity test

IEC60601 test level 

Electromagnetic environment - guidance

Conducted RF
IEC61000-4-6

Radiated RF
IEC61000-4-3

3Vrms
150 kHz to 80MHz

3V/m
80MHz to 2.5 GHz

Compliance level 

3 Vrms

3 V/m

Portable and mobile RF communications equipment should be used 
no closer to any part of the ENDO-MATE DT, including cables, than 
the recommended separation distance calculated from the equation 
applicable to the frequency of the transmitter.

Recommended separation distance

d

 = 1.2   

P

d

 = 1.2   

P

  80MHz to 800MHz

d

 = 2.3   

P

  800MHz to 2.5GHz

Where P is the maximum output power rating of the transmitter in 
watts (W) according to the transmitter manufacturer and d is the 
recommended separation distance in metres (m).

Field strengths from fixed RF transmitters as determined by an 
electromagnetic site survey, should be less than the compliance 
level in each frequency range.

Interference may occur in the vicinity of equipment 
marked with the following symbol:

Complies with

Maximum length

Cables and accessories

RF emissions, CISPR11,

Harmonic emissions,

Voltage fluctuations/ flicker emission

Electrostatic discharge (ESD)

Surge

Voltage dips, short interruptions and voltage variations on power supply input lines

Power frequency (50/60Hz) magnetic field

Conducted RF

Radiated RF

1.5 m

2.5 m

Plug side : 1.8m

Unit side : 1.8m

Motor Handpiece cord

Foot Controler cord

AC Adapter

Class B/ Group 1

IEC61000-3-2, Class A

IEC61000-3-3

IEC61000-4-2

IEC61000-4-5

IEC61000-4-11

IEC61000-4-8

IEC61000-4-6

IEC61000-4-3

150kHz to 80MHz

d

 =1.2    

P

Recommended separation distances between portable and mobile RF communications equipment and the ENDO-MATE DT

The ENDO-MATE DT is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the 
ENDO-MATE DT can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment 
(transmitters) and the ENDO-MATE DT as recommended below, according to the maximum output power of the communications equipment.

For transmitters rated at a maximum output power not listed above, the recommended separation distance 

d

  in meters (m) can be estimated using the equation 

applicable to the frequency of the transmitter, where 

P

  is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.

Rated maximum output power of transmitter

W

800MHz to 2.5GHz

d

 =2.3    

P

80MHz to 800MHz

d

 =1.2    

P

NOTE 1

  

At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.

NOTE 2

 

These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.

Separation distance according to frequency of transmitter

m

0.01

0.1

1

10

100

0.12
0.38

1.2
3.8

12

0.12
0.38

1.2
3.8

12

0.23
0.73

2.3
7.3

23

Summary of Contents for ENDO-MATE DT

Page 1: ...OPERATION MANUAL Please read this Operation Manual carefully before use and file for future reference OM E0202E 002 Motor Handpiece System for Root Canal Enlargement...

Page 2: ...this Operation Manual for future reference Read these safety cautions thoroughly before use and operate the product properly These indicators are to allow you to use the product safely and prevent da...

Page 3: ...ged deformed or non ISO conforming file Using such a file may result in personal injury due to its unexpected breaking or flying off during rotation Do not use or leave the product in a high temperatu...

Page 4: ...though it is not used It is recommended to recharge the batteries just before use When the motor handpiece stops automatically by detecting a low battery voltage turning on the power again after a whi...

Page 5: ...luding motor handpiece cord Fig 1 2 3 8 10 9 1 12 4 5 6 7 11 Option Control Unit Operation Panel See Fig 2 Operation Panel and LCD Panel LCD Panel See Fig 2 Operation Panel and LCD Panel Error codes a...

Page 6: ...pending on the rotation speed Operation Panel and LCD Panel Fig 2 LCD Panel This indicates that all displays are lighted TORQUE Key The torque limit values can be changed If the speed is to exceed the...

Page 7: ...available at 1 1 4 1 10 1 16 1 and 20 1 REVERSE Key The rotational direction of the file can be changed by this key It can also be changed while the file is in motion The alarm sounds during reverse...

Page 8: ...on of the power LCD The mark of the present remaining amount of the batteries is displayed regardless of the ON or OFF position of the power The mark will be animated when the batteries are being char...

Page 9: ...he mark of the plug cast down Fig 5 You do not have to connect the Foot Control if you intend to use this product without the Foot Control 4 HOW TO CONNECT EACH PART To remove the plug hold the plug r...

Page 10: ...imes The temperature of the batteries is measured during charging Proper charging cannot be preformed if the charger is placed in an environment which is subject to sharp temperature change next to wi...

Page 11: ...t gear ratio is selected according to the contra angle head in use an adequate rotation speed and torque limit value can be set 4 Select the rotational direction by pressing the REVERSE key The key ch...

Page 12: ...tion 5 Operation 1 Start the motor handpiece to pressing the ON OFF button briefly Press the button again may stop the handpiece Alternate operation Keep pressing ON OFF button more than a second the...

Page 13: ...or more than one second If the program number is changed by the PROGRAM key without pressing the MEMORIZE key the initially memorized preset values remain Cancel function NOTICE 6 Completion of Medica...

Page 14: ...play at the time of charging At the same time the CHARGE lamp flashes slowly 5 Leave it as it is The batteries are discharged and recharged automatically 6 Hold down the POWER key for a while when you...

Page 15: ...ad for 2 3 seconds Fig 12 2 Cleaning Motor Handpiece When the motor handpiece becomes dirty wipe it off with a cotton cloth moistened with rubbing alcohol 7 CLEANING Do not lubricate the motor section...

Page 16: ...out the old batteries and pull out the cord holding it at the connector Fig 14 4 Insert the connector of the cord of new batteries into the connector in accordance with the polarity indication label...

Page 17: ...on of the circuit may be suspected Contact your dealer Malfunction of circuit The motor handpiece is locked at the time of the auto reverse mode Malfunction of circuit High load was continuously appli...

Page 18: ...batteries is high An error code is displayed Check the connection Check the connection Charge the batteries or use the AC Adaptor Insert batteries or use the AC Adaptor Contact your dealer Check the c...

Page 19: ...t is rotated by the Foot Control Option Clean or replace the contra angle head Contact your dealer Check the ON OFF button Contact your dealer Check the Foot Control Option Remove the foot control Opt...

Page 20: ...tronic equipment The ENDO MATE DT is suitable for use in all establishments including domestic establishments and those directly connected to the public low voltage power supply network that supply ne...

Page 21: ...electromagnetic immunity The ENDO MATE DT is intended for use in the electromagnetic environment specified below The customer or the user of the ENDO MATE DT should assure that it is used in such an e...

Page 22: ...vent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment transmitters and the ENDO MATE DT as recommended below according to the maxi...

Page 23: ...Specifications are subject to change without notice 2014 11 20 02 M The EU directive 93 42 EEC was applied in the design and production of this medical device...

Reviews: