NSK ENDO-MATE DT Operation Manual Download Page 18

16

17

When trouble is found, check the following again before contacting your dealer. If none of these is applicable or the trouble 
is not remedied even after action has been taken, a failure of this product is suspected. Contact your dealer.

<Control Unit, AC Adaptor>

Problem

Cause

Solution

11. TROUBLESHOOTING

If the motor handpiece stops due to an abnormality such as a malfunction, overload, break or wrong use, it automatically 
checks the state of the control unit and detects the cause of the abnormality and displays an error code on the LCD panel. If 
an error code is displayed, turn on the power again and check whether the same error code is displayed. If the same error 
code is displayed, take action by referring to the instructions provided in the “Check/Remedy” column in the following table.

10. ERROR CODE

If batteries are replaced while the AC Adaptor is connected, an incorrect error code may be displayed.  

NOTICE

Check/Remedy

Cause

Error

Error code

Contact your dealer.

Remove load

Contact your dealer.

Leave it as it is until it cools down.

Contact your dealer.

Remove load

Put the batteries into the battery 
chamber, or replace with new 
batteries.

Contact your dealer.

Use within the range of working 
temperature, or replace the batteries.

Replace the motor handpiece or 
contra angle head.

Replace the batteries. If the heat 
generates from the new set of 
batteries, malfunction of the circuit 
may be suspected. Contact your dealer.

Malfunction of circuit

The motor handpiece is 
locked. (at the time of the 
auto reverse mode)

Malfunction of circuit

High load was continuously 
applied to the motor 
handpiece for a relatively 
long time.

Malfunction of circuit

The motor handpiece is 
locked. (at the time of startup)

The voltage of batteries is too 
low (The life of batteries or 
not iserted.)

The voltage of batteries is too 
high. (Malfunction of circuit)

Beyond the range of working 
temperature or break in the 
thermistor in the battery 
section.

The operating life of the 
motor handpiece or contra 
angle has expired.

The batteries generate 
abnormal heat.

Self-Check error

Ovecurrent

Overvoltage

Overheating of motor

Error of the brake 
circuit

The axle locks up

Low voltage of 
batteries

High voltage of 
batteries 

Beyond the range of 
working temperature 

Beyond the upper limit

Below the lower limit

Abnormal heat 
generation from 
batteries

During 
rotation of 
the motor 
handpiece

At the time 
of charging

At the time 
of calibration

Other

<Motor Handpiece>

Problem

Cause

Solution

5) Close the battery cover. Make sure to charge the batteries before use.

The used nickel metal hydride batteries are recyclable, but their disposal may sometimes not be permitted by the 
country. Return them to your dealer.

NOTICE

The power is not turned on. 

AC IN lamp does not light.

The battery charger does 
not work. 
(The CHARGE lamp does 
not light.)

The AC Adaptor is not connected.

The plug of the AC Adaptor is not inserted 
into the outlet, or there is no electricity in 
the outlet.

The batteries are empty.

No batteries are inserted.

The fuse has burnt.

The AC Adaptor is not connected.

The plug of the AC Adaptor is not inserted 
into the outlet, or there is no electricity in 
the outlet.

The fuse has burnt.

The AC Adaptor’s fuse has burnt.

No batteries are inserted.

The batteries are fully charged or in a state 
near full charge.

The temperature of batteries is low.

The temperature of batteries is high.

An error code is displayed.

Check the connection.

Check the connection.

Charge the batteries or use the AC Adaptor.

Insert batteries, or use the AC Adaptor.

Contact your dealer.

Check the connection.

Check the connection.

Contact your dealer.

Contact your dealer.

Insert batteries.

No problem

If the temperature of batteries is less than 
0°C (32°F), the batteries are not 
rechargeable. Charge the batteries in a 
warm room. (Be careful about moisture 
condensation.)

It is normal that the batteries become a 
little bit warm right after charging. If the 
batteries are hot under normal operating 
conditions, not right after charging, there 
may be an abnormality. Contact your dealer.

Refer to 

P.16

  10. ERROR CODE.

The motor handpiece does 
not rotate.

The motor handpiece does 
not rotate. (The error code 
“               ” is displayed.)

The motor handpiece cord is not connected.

The Foot Control (Option) is not connected.

There is a break in the motor handpiece or 
the motor handpiece cord.

The contra angle head is clogged.

There is a short circuit inside the motor 
handpiece or the motor handpiece cord.

Check the connection.

Check the connection.

Contact your dealer.

Clean or replace the contra angle head.

Contact your dealer.

Summary of Contents for ENDO-MATE DT

Page 1: ...OPERATION MANUAL Please read this Operation Manual carefully before use and file for future reference OM E0202E 002 Motor Handpiece System for Root Canal Enlargement...

Page 2: ...this Operation Manual for future reference Read these safety cautions thoroughly before use and operate the product properly These indicators are to allow you to use the product safely and prevent da...

Page 3: ...ged deformed or non ISO conforming file Using such a file may result in personal injury due to its unexpected breaking or flying off during rotation Do not use or leave the product in a high temperatu...

Page 4: ...though it is not used It is recommended to recharge the batteries just before use When the motor handpiece stops automatically by detecting a low battery voltage turning on the power again after a whi...

Page 5: ...luding motor handpiece cord Fig 1 2 3 8 10 9 1 12 4 5 6 7 11 Option Control Unit Operation Panel See Fig 2 Operation Panel and LCD Panel LCD Panel See Fig 2 Operation Panel and LCD Panel Error codes a...

Page 6: ...pending on the rotation speed Operation Panel and LCD Panel Fig 2 LCD Panel This indicates that all displays are lighted TORQUE Key The torque limit values can be changed If the speed is to exceed the...

Page 7: ...available at 1 1 4 1 10 1 16 1 and 20 1 REVERSE Key The rotational direction of the file can be changed by this key It can also be changed while the file is in motion The alarm sounds during reverse...

Page 8: ...on of the power LCD The mark of the present remaining amount of the batteries is displayed regardless of the ON or OFF position of the power The mark will be animated when the batteries are being char...

Page 9: ...he mark of the plug cast down Fig 5 You do not have to connect the Foot Control if you intend to use this product without the Foot Control 4 HOW TO CONNECT EACH PART To remove the plug hold the plug r...

Page 10: ...imes The temperature of the batteries is measured during charging Proper charging cannot be preformed if the charger is placed in an environment which is subject to sharp temperature change next to wi...

Page 11: ...t gear ratio is selected according to the contra angle head in use an adequate rotation speed and torque limit value can be set 4 Select the rotational direction by pressing the REVERSE key The key ch...

Page 12: ...tion 5 Operation 1 Start the motor handpiece to pressing the ON OFF button briefly Press the button again may stop the handpiece Alternate operation Keep pressing ON OFF button more than a second the...

Page 13: ...or more than one second If the program number is changed by the PROGRAM key without pressing the MEMORIZE key the initially memorized preset values remain Cancel function NOTICE 6 Completion of Medica...

Page 14: ...play at the time of charging At the same time the CHARGE lamp flashes slowly 5 Leave it as it is The batteries are discharged and recharged automatically 6 Hold down the POWER key for a while when you...

Page 15: ...ad for 2 3 seconds Fig 12 2 Cleaning Motor Handpiece When the motor handpiece becomes dirty wipe it off with a cotton cloth moistened with rubbing alcohol 7 CLEANING Do not lubricate the motor section...

Page 16: ...out the old batteries and pull out the cord holding it at the connector Fig 14 4 Insert the connector of the cord of new batteries into the connector in accordance with the polarity indication label...

Page 17: ...on of the circuit may be suspected Contact your dealer Malfunction of circuit The motor handpiece is locked at the time of the auto reverse mode Malfunction of circuit High load was continuously appli...

Page 18: ...batteries is high An error code is displayed Check the connection Check the connection Charge the batteries or use the AC Adaptor Insert batteries or use the AC Adaptor Contact your dealer Check the c...

Page 19: ...t is rotated by the Foot Control Option Clean or replace the contra angle head Contact your dealer Check the ON OFF button Contact your dealer Check the Foot Control Option Remove the foot control Opt...

Page 20: ...tronic equipment The ENDO MATE DT is suitable for use in all establishments including domestic establishments and those directly connected to the public low voltage power supply network that supply ne...

Page 21: ...electromagnetic immunity The ENDO MATE DT is intended for use in the electromagnetic environment specified below The customer or the user of the ENDO MATE DT should assure that it is used in such an e...

Page 22: ...vent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment transmitters and the ENDO MATE DT as recommended below according to the maxi...

Page 23: ...Specifications are subject to change without notice 2014 11 20 02 M The EU directive 93 42 EEC was applied in the design and production of this medical device...

Reviews: