4
VCM-3 Homepage Commissioning Procedure
After the visible inspection check the on-site commissioning procedure shall be continued using the
functionality of the VCM-3 Homepage.
Via the Homepage the following steps shall be completed:
1. Check the VCM-3 firmware version
2. Check the Operational Status
3. Check the sensor bias voltages
4. Perform a Tap test of all accelerometers in order to check the correct mounting of the
accelerometers
5. Check the rotational speed sensor signal using a piece of metal
6. Configure the VCM-3 with the correct IP addresses and enable services
7. Save the commissioning report
8. Reboot the VCM-3 after completion
9. Inspect the VCM-3 front plate after reboot
4.1
Log into the VCM-3 Homepage
How to connect a laptop and perform the login.
1
Change the laptop network settings to match the ip range of the VCM-3. Factory default network
settings of VCM-3 are as follows:
•
Netmask (subnet mask): 255.255.255.0
•
VCM-3 Ipv4 address: 192.168.2.202
NOTE!
Brüel & Kjær Vibro recommends that to set the laptop to Ipv4 address
192.168.2.199, Netmask 255.255.255.0
2
Connect the laptop to one of the RJ45 ports of the VCM-3 unit using a LAN cable.
RJ45 cable connector.
3
Write the VCM-3 IP address in the browser. Ignore the security notifications and proceed to the
VCM-3 login page.
Table continued on next page…