1 DESCRIPTION AND OPERATION OF THE DEVICE AND ITS COMPONENTS

1.1 The purpose of the device

Vibration Meter Pen NVS P1 is used to study the vibration level of various mechanisms.

The vibration level can be judged by acceleration, speed and degree of displacement. The

Vibration Meter Pen is equipped with a built-in high-sensitivity vibration sensor in the form

of a probe. The Vibration Meter Pen transducer transmits analog information about

vibrations to the device; the Vibration Meter Pen electronics processes the received data and

displays them on the LCD display.

The Vibration Meter Pen is used to measure vibration according to the following

parameters:

− vibration velocity, mm / s;

− vibration displacement, mm;

− ratio of peak and root-mean-square values of vibration;

− vibration acceleration, mm / s

2

.

The device is equipped with an auto-off function - after 20 seconds of inactivity, the

Vibration Meter Pen will turn off automatically.

The device is powered by a built-in battery.

The instrument is widely used in petrochemical and machinery industries, metallurgy,

vehicles, etc.

1.2 Technical characteristics of the device

The main characteristics of the device are presented in table. 1.1.

Table 1.1 - Main characteristics of the device

Vibration acceleration (amplitude):

0.01 ÷ 50.0 m / s

2

Vibration speed:

up to 35 mm / s (160 Hz)

Vibration displacement (span):

up to 563 mm / s (10 Hz)

Measured frequency range:

up to 100 microns (160 Hz)

Accuracy:

up to 5000 microns (10 Hz)

Duration of continuous work, h

10 ÷ 1000 Hz

Charge duration, h

± 5% (10 ÷ 1000 Hz)

Connector type

up to 8

Overall dimensions, mm

2.5

Degree of protection

USB - micro B USB 1.0 Receptacle Connector

Weight, g, no more

157 × 22 × 14

Mean time between failures - not less than 2.400 hours.

Average full service life is 5 years.

1.3 Standard delivery set

− Measuring unit

1 pc.

− Charger

1 pc.

− Packing

1 pc.

− Operating manual

1 pc.

− Calibration certificate

1 pc.

Summary of Contents for NVS P1

Page 1: ...NVS P3...

Page 2: ...rking and sealing 5 1 6 Packing 5 2 INTENDED USE 6 2 1 Operational limitations 6 2 2 Preparing the device for use 6 2 3 Using the device 6 2 3 1 Turning on 6 2 3 2 Measurement modes 6 2 3 3 Measuremen...

Page 3: ...necessary to read this manual as the device operation must be carried out by persons who are familiar with the operation principle and the device design Correct and effective use of the hardness test...

Page 4: ...function after 20 seconds of inactivity the Vibration Meter Pen will turn off automatically The device is powered by a built in battery The instrument is widely used in petrochemical and machinery in...

Page 5: ...uses the LCD 2 3 4 5 1 1 vibration sensor 2 Vibration Meter Pen body 3 Modes key 4 LCD display 5 key for switching on and measuring Fig 1 1 Vibration Meter Pen NVS P1 1 5 Marking and sealing The manuf...

Page 6: ...ed on the right side of the display you need to connect the device to a PC using a USB cable or to a charger Note When using the Vibration Meter Pen for the first time it is necessary to conduct an ex...

Page 7: ...or with the object press and hold the key the device will start measuring vibration The measurement result is displayed on the LCD During of measuring vibration the device makes cyclic measurements at...

Page 8: ...ntees the conformity of the device to the requirements of the technical conditions under the user s compliance with the conditions of transportation storage and operation and timely maintenance at the...

Page 9: ...r short term storage and between use the device should be stored in the intended packaging The storage area should be free from vapors of aggressive substances acids alkalis and direct sunlight The de...

Reviews: