Section 10
Maintenance
36
Model 520A Service Manual
5693-90-01
PRELIMINARY
Cleaning and Sterilization
10.3
Model 520A Monitor
10.3.1
• Turn the monitor off and unplug from the AC (Mains) line before cleaning.
• The monitor can be cleaned and disinfected with solutions such as a 70%
isopropyl alcohol, 2% glutheralhyde, or 10% bleach solution. Then wipe down
with a water-dampened clean cloth to rinse. Dry before use.
• Do not immerse the monitor.
• Do not attempt to sterilize the monitor.
Finger Sensor
10.3.2
• The sensor can be cleaned and disinfected with solutions such as a 70%
isopropyl alcohol, 2% glutheralhyde, or 10% bleach solution. Then wipe down
with a water-dampened clean cloth to rinse. Dry before use.
• Make certain that the finger sensor windows are clean and dry before reuse.
• Do not immerse the finger sensor.
• Do not attempt to sterilize the finger sensor.
• After cleaning the finger sensor, verify that the sensor is physically intact, with
no broken or frayed wires or damaged parts. Make certain that the connectors
are clean and dry, with no signs of contamination or corrosion. Do not use a
broken or damaged sensor or one with wet, contaminated or corroded
connectors.
Y-SENSOR™ and Y-STRIP™ Taping System
10.3.3
• Do not immerse connector on the Y-Sensor.
• The Y-Sensor may be immersed—up to, but not including, the connector, in a
2% glutheralhyde solution, or 10% bleach solution. Refer to manufacturer’s
instructions and standard hospital protocols to determine recommended times
for disinfection and sterilization.
• Rinse thoroughly with water and dry before use (do not rinse the connector).
• Do not attempt to sterilize Y-Sensor except as stated above.
• After cleaning or sterilizing the Y-Sensor, verify that the sensor is physically
intact, with no broken or frayed wires or damaged parts. Make certain that the
connectors are clean and dry, with no signs of contamination or corrosion. Do
not use a broken or damaged sensor or one with wet, contaminated, or corroded
connectors.
• Treat Y-Strip Tapes and foam wraps in accordance with hospital protocol for
single-patient use items.
Summary of Contents for OXYPLETH 520A
Page 10: ...Section List of Tables ix Model 520A Service Manual 5693 90 01 PRELIMINARY...
Page 104: ......
Page 105: ......
Page 106: ......
Page 107: ......
Page 108: ......
Page 109: ......
Page 110: ......
Page 111: ......
Page 112: ......
Page 113: ......
Page 114: ......
Page 115: ......
Page 116: ......
Page 117: ......
Page 118: ......
Page 119: ......
Page 120: ......
Page 121: ......
Page 122: ......
Page 123: ......
Page 124: ......
Page 125: ......
Page 126: ......
Page 127: ......
Page 128: ......
Page 129: ......
Page 130: ......
Page 131: ......
Page 132: ......
Page 133: ......