DOC 225-HMP60x UM 20210202
225-HMP60x
User Manual
225-HMP60A (0-1 V outputs)
225-HMP60C (0-5 V outputs)
Temperature and Relative Humidity Sensor
Phone
(530) 823-7185
Email
[email protected]
Website
www.novalynx.com
Page 1: ...225 HMP60x UM 20210202 225 HMP60x User Manual 225 HMP60A 0 1 V outputs 225 HMP60C 0 5 V outputs Temperature and Relative Humidity Sensor Phone 530 823 7185 Email nova novalynx com Website www novalynx...
Page 2: ...e items If packing facilities are not available take the equipment to the nearest Post Office UPS or other freight service and obtain assistance with the packaging Please write the RA on the outside o...
Page 3: ...Corporation 225 HMP60x Page 3 February 2021 CONTENTS 1 FORWARD 4 2 INTRODUCTION 4 3 WIRE INSTALLATION TABLE 4 4 SPECIFICATIONS 5 5 INSTALLATION 6 6 MAINTENANCE 6 7 CALIBRATION TEST KIT 7 8 TROUBLESHO...
Page 4: ...or range is 0 to 100 relative humidity Temperature is measured by a 1000 ohm RTD element The temperature range is 40 to 60 C 40 to 140 F Model 225 HMP60A features two 0 1 V outputs Model 225 HMP60C fe...
Page 5: ...HMP60A RL min 10 k HMP60C RL min 50 k Accuracy at 20 C 68 F 0 2 of FS Temperature dependence 0 01 of FS C 0 006 of FS F Power consumption 1 mA average max peak 5 mA Operating voltage HMP60A 5 28 VDC H...
Page 6: ...e usually positioned at shelter height or about four feet above ground level on the north side of the mast Temperature readings can be affected by heat reflected from the ground so try to avoid instal...
Page 7: ...or 7 CALIBRATION TEST KIT The NovaLynx 220 HMK15 Humidity Calibrator sold separately is available for checking the calibration of electronic RH sensors by the user The calibrator includes two chambers...