The environmental temperature range for body temperature measuring is 0~55°C, the
minimum measuring time is 1s, and the measuring interval is 1s.
Warning
Please measure the body temperature in the specified environmental temperature
range, or else it may be dangerous.
11.5
Temp Display
The monitor can display the body temperature of two channels (T1 and T2) and the alarm
limits, difference between the two temperature (TD) and temperature units. Select Temp
parameter area and open the [Temp Setup] menu.
Temp display area is shown below:
Fig. 12-1 TEMP Parameter Area
11.6
Selecting Resistance Type of Temp Probe
Select the appropriate [Temp Sensor] value according to the actual resistance of the
temperature probe being used. The specific setting method follows:
Select
[Menu] smart hotkey → [Main Menu];
Select [Factory Maintenance] → Enter and confirm the password → [Factory
Maintenance] menu;
In [Factory Maintenance] menu, set [Temp Sensor] to
[10K], probe resistance is 10k
Ω
;
[2K], probe resistance is 2.25kΩ.
Warning