In PC-ACV mode, you need to set the following basic ventilation parameters:
1.
[Pmax]: Can prevent high airway pressure caused by
patient
2.
[PEEP]: Positive end-expiratory pressure
3.
[Pinsp]: Inspiration pressure
4.
[I:E]: Inspiration time
5.
[Rate]: Breathing frequency
6.
[Tslope]: Time of pressure rising
7.
[Trigger]: Inspiration trigger level
8.
[O
2
%]: Oxygen concentration
In PC-ACV mode, you can set the following sigh function parameters as required:
1.
[
Sigh
]: Switch for turning on sigh function
2.
[
Interval
]: Number of sigh cycles
3.
[
Delta Pinsp
]:
Pinsp added in sigh cycle
6.5.5 PVC-ACV
PVC-A/C is pressure- regulated volume /control ventilation mode. It implements delivering set
tidal volume by the way of pressure control ventilation. In PRVC, a relatively low pressure
level is held as much as possible during the inspiratory phase, and the gas volume delivered is
Summary of Contents for Crius V6
Page 1: ...1 182 Crius V6 Ventilator Operator s Manual...
Page 15: ...15 182 FOR YOUR NOTES...
Page 26: ...26 182 FOR YOUR NOTES...
Page 28: ...28 182 2 2 Equipment Appearance 2 2 1 Front View 1 13 9 10 12 11 8 7 6 5 2 4 3...
Page 32: ...32 182 FOR YOUR NOTES...
Page 44: ...44 182 FOR YOUR NOTES...
Page 56: ...56 182 FOR YOUR NOTES...
Page 88: ...88 182 FOR YOUR NOTES...
Page 100: ...100 182 FOR YOUR NOTES...
Page 125: ...125 182 FOR YOUR NOTES...
Page 149: ...149 182 FOR YOUR NOTES...
Page 156: ...156 182 FOR YOUR NOTES...