15.11 Factory Defaults
This chapter lists the most important factory default settings which are not
user-adjustable. When necessary, you can restore the factory default settings.
15.11.1 Ventilation Mode
Ventilation
mode
Parameter
Defaults setting
Adu
Ped
Neo
VC-ACV
Pmax
100mbar
40mbar
40mbar
PEEP
3mbar
3mbar
3mbar
Vt
536mL
150mL
150mL
Tinsp
2.0s
1.0s
0.6s
Rate
20bpm
20bpm
30bpm
Tpause
OFF
OFF
OFF
Trigger
2.0L/min
1.0L/min
0.2L/min
Sigh
OFF
OFF
OFF
ATC
Disable ATC
Disable ATC
Disable ATC
Compliance Compensate
OFF
OFF
OFF
Exp.Speed
Medium
Medium
Medium
PC-ACV
Pmax
100mbar
40mbar
40mbar
PEEP
3mbar
3mbar
3mbar
Pinsp
15mbar
15mbar
15mbar
Tinsp
2.0s
1.0s
0.6s
Rate
20bpm
20bpm
30bpm
Tslope
50%
50%
50%
Trigger
2.0L/min
1.0L/min
OFF
Sigh
OFF
OFF
OFF
ATC
Disable ATC
Disable ATC
Disable ATC
Compliance Compensate
OFF
OFF
OFF
Exp.Speed
Medium
Medium
Medium
PVC-ACV
Pmax
100mbar
40mbar
40mbar
PEEP
3mbar
3mbar
3mbar
Pinsp
15mbar
15mbar
15mbar
Tinsp
2.0s
1.0s
0.6s
Rate
20bpm
20bpm
30bpm
Tslope
50%
50%
50%
Trigger
2.0L/min
1.0L/min
OFF
Sigh
OFF
OFF
OFF
ATC
Disable ATC
Disable ATC
Disable ATC
Compliance Compensate
OFF
OFF
OFF
Summary of Contents for Crius V6
Page 1: ...1 182 Crius V6 Ventilator Operator s Manual...
Page 15: ...15 182 FOR YOUR NOTES...
Page 26: ...26 182 FOR YOUR NOTES...
Page 28: ...28 182 2 2 Equipment Appearance 2 2 1 Front View 1 13 9 10 12 11 8 7 6 5 2 4 3...
Page 32: ...32 182 FOR YOUR NOTES...
Page 44: ...44 182 FOR YOUR NOTES...
Page 56: ...56 182 FOR YOUR NOTES...
Page 88: ...88 182 FOR YOUR NOTES...
Page 100: ...100 182 FOR YOUR NOTES...
Page 125: ...125 182 FOR YOUR NOTES...
Page 149: ...149 182 FOR YOUR NOTES...
Page 156: ...156 182 FOR YOUR NOTES...