Under normal operating conditions for medical oxygen delivery equipment the sensors are warranted
to be free of defects in materials and workmanship for the period specified period above provided the
sensor is properly installed and operated. The sole remedy for a sensors determined to be defective by
Analytical Industries Inc. is limited to replacing the defective sensors. Analytical Industries Inc. shall
not be liable for buyer’s negligence, misapplication, alteration, abuse or accident.
15.10 SpO2 Specifications (optional)
Standards compliant
ISO 80601-2-61
Display range
0%~100%
SpO2 display
resolution
1%
SpO2 checking
accuracy
2% (70%~100%); not define when lower than 70%
SpO2 alarm preset
limits
Upper alarm limit
1%~100%
Lower alarm limit
0%~99%
SpO2 alarm preset
accuracy
±1%
SpO2 alerting signal
generates a delay
No delay
SpO2 value refresh
period
1s/time
SpO2 value refresh
delay
<
10s
Average period
Low sensitivity
7
~
8s
Intermediate sensitivity
4
~
6s
Advanced sensitivity
2
~
3s
Alarm condition delay
period
Low sensitivity
<
8s
Intermediate sensitivity
<
6s
Advanced sensitivity
<
3s
Alarm sign generates
delay period
0s
PR
Measuring range
30~254bpm
Resolution
1%
Accuracy
±2% or ±2bpm
Summary of Contents for Crius V6
Page 1: ...1 182 Crius V6 Ventilator Operator s Manual...
Page 15: ...15 182 FOR YOUR NOTES...
Page 26: ...26 182 FOR YOUR NOTES...
Page 28: ...28 182 2 2 Equipment Appearance 2 2 1 Front View 1 13 9 10 12 11 8 7 6 5 2 4 3...
Page 32: ...32 182 FOR YOUR NOTES...
Page 44: ...44 182 FOR YOUR NOTES...
Page 56: ...56 182 FOR YOUR NOTES...
Page 88: ...88 182 FOR YOUR NOTES...
Page 100: ...100 182 FOR YOUR NOTES...
Page 125: ...125 182 FOR YOUR NOTES...
Page 149: ...149 182 FOR YOUR NOTES...
Page 156: ...156 182 FOR YOUR NOTES...