Chapter 1 -About the DR400
The NorthEast Monitoring DR400 patch style recorder can be used as either a Holter mon-
itor and/or a looping Event recorder. Data obtained by monitoring is not analyzed at the
time of recording. After recording is complete, Holter data is uploaded to a compatible ver-
sion of NorthEast Monitoring’s Holter Analysis software. The DR400 recorder is compati-
ble with 5.2 or newer versions of HE/LX Analysis software. For Event, data is sent
wirelessly to a location where versions 2.11 or newer of the LX Event utility is used to
review and report on the recording.
The DR400 Recorder is not intended to replace real-time telemetry monitoring for
patients suspected of having life-threatening arrhythmias.
The DR400 Recorder is not for In Vitro diagnostic use.
Specifications
Physical Specifications
The DR400 recorder meets the following physical specifications:
• Size: 6.8 cm (length) x 3.9 cm (width) x 1.3 cm (depth)
• Weight: 33 grams (1.2 oz.) with battery
Electrical Specifications
The DR400 recorder electrical specifications are:
•
Recording bandwidth: 0.05 to 70 hertz in 180 samples/sec. mode*
•
Operation duty cycle: Continuous
•
Data storage format: Sample difference
•
Pacemaker sensitivity: 2 millivolts
•
Pacemaker pulse duration: 150 to 2,500 microseconds
•
Resettable fuses: 0.5 amp
* When measuring millivolts, results should be within 5% of expected results.