NorthEast Monitoring, Inc. NEMM048-Rev-G
Page 14 of 21
DR400 User Manual
PCPatch & Wireless: ETel – Mobile Cardiac Telemetry (MCT) monitoring
ETel – Mobile Cardiac
Telemetry (MCT) monitoring
NorthEast Monitoring’s ETel utility must be
installed on your PC for Mobile Cardiac Telemetry
(MCT). MCT is a feature that can be added to a
DR400 in Event recording mode and allows for
bidirectional communication between PC and the
DR400 without patient interaction.
MCT allows you to request continuous recording
for a period of time within the past 24 hours and
also allows you to change any or all of the Auto
Detect Settings during the event procedure.
The ETel Utility
The ETel utility is used to send MCT messages to
the DR400 during the procedure. The screen below
shows a request that would be used to obtain ECG
data. ECG data requests must be within the last 24
hours. Up to 10 minutes at a time can be requested.
Multiple requests can be sent during a single proce-
dure, however, only one request at a time can be
queued for a single DR400.
Additionally,
Auto-event
settings can
be updated
during the
procedure.
The values
for the next
setting are put
in the New Setting Value field.
Once a request is Sent, it will be
queued in the window on the
right. For each DR400, only one
request can be sent at a time.
The request will be removed
from the list once it has been received by the
DR400.
If you need to delete a command before it is sent,
you can click on it while on the list and confirm that
you want it removed.