Preferences: Research/Small Subject Scan
Prior to starting a Research or Small Subject
1
study, the scan parameters should be defined. The desired pre-
cision and detail of a scan, the scan length, and the time to scan completion influence the scan parameter selec-
tions.
In order to maintain precision and accuracy, always use the same scan parameters for serial scans.
The following figure shows the default Research/Small Subject Scan preferences.
Figure 3-10:
The default Research/Small Subject Scan Parameters
Note:
WHAT TO DO: To restore the
default
Scan parameters, click
and
. This will restore the defaults for
all
scan types.
Note:
WHAT TO DO: To
change
the default Scan parameters, access the Scanner Preferences win-
dow from the main Database Navigator window as described next.
If it is not already open, click on Edit > Scanner Preferences > Scanning tab.
Locate the Research/Small Subject scan parameters in the window (shown above).
Select and change any of the scan parameters, such as resolution, scan width, or speed.
Click
and
at the bottom of the window. These settings now become the
default settings until changed again.
PARAMETER DESCRIPTIONS:
Resolution: values refer to point resolution (pixel size) x line spacing
Scan Speed: entered in mm/sec
Minimum for
Scout
and
Measure
scans is 1.0mm/sec
Maximum for
Scout
and
Measure
scans is 260.0mm/sec
1
Research and/or Small Subject are not available on some models.
Operator's Guide
437D140 Rev. I
3-28
Installation and Setup
Summary of Contents for Elite 437A150
Page 1: ...DXA Bone Densitometer Operator s Guide Elite Model 437A150 Operator s Guide 437D140 Rev I...
Page 22: ...This page intentionally left blank Operator s Guide 437D140 Rev I 1 8...
Page 56: ...This page intentionally left blank Operator s Guide 437D140 Rev I 2 34...
Page 66: ...11 Click Operator s Guide 437D140 Rev I 3 10 Installation and Setup...
Page 108: ...This page intentionally left blank Operator s Guide 437D140 Rev I 3 52...
Page 124: ...Sample Printed QA Results Report Operator s Guide 437D140 Rev I 4 16 Basic Operation...
Page 137: ...Operator s Guide 437D140 Rev I Basic Operation 4 29...
Page 166: ...This page intentionally left blank Operator s Guide 437D140 Rev I 5 24...
Page 198: ...Operator s Guide 437D140 Rev I 7 20 Scanning AP Spine...
Page 206: ...This page intentionally left blank Operator s Guide 437D140 Rev I 7 28...
Page 238: ...Figure 8 6 Page 2 of the Bone Exam Report Operator s Guide 437D140 Rev I 8 32 Scanning Hip...
Page 290: ...This page intentionally left blank Operator s Guide 437D140 Rev I 10 30...
Page 326: ...Operator s Guide 437D140 Rev I 11 36 Additional Techniques...
Page 412: ...This page intentionally left blank Operator s Guide 437D140 Rev I 14 16...
Page 428: ...This page intentionally left blank Operator s Guide 437D140 Rev I 16 10...
Page 446: ...Label Locations continued Operator s Guide 437D140 Rev I 17 18 Technical Reference...
Page 448: ...Scanner Labels continued Operator s Guide 437D140 Rev I 17 20 Technical Reference...
Page 449: ...Operator s Guide 437D140 Rev I Technical Reference 17 21 Scanner Labels continued...
Page 459: ...Operator s Guide 437D140 Rev I Reference Data Sets 18 9...
Page 466: ...Operator s Guide 437D140 Rev I This page intentionally left blank...
Page 484: ...Operator s Guide 437D140 Rev I This page intentionally left blank...