Operator's Guide
437D140 Rev. I
Quick Reference - QA Calibration
Daily Calibrations should be performed prior to patient scanning to ensure quality bone density estimates. If a
successful calibration has not been completed within the last 14 days, scanning will be disabled.
Check the Calibration Standard for bent corners, damaged plastic, or loose parts (when shaken). If dam-
aged, contact Norland Customer Service.
Click on Calibration > Begin QA in the DXA Database Navigator window. Click on
in the Cal-
ibration tab window. The system will prompt the operator to place the calibration standard on the scanner and
mark point A.
Place the standard on the patient surface in the proper orientation as shown. Align it with the corresponding
marks on the table surface.
Switch ON the positioning laser.
Caution:
Do not stare into the beam.
Use the arrow buttons to move the scanner arm so the laser positioning dot is on the ‘+’ point on the Cal-
ibration Standard Plexiglas surface identified by the inscribed letter “A” and press the MARK button.
Use the arrow buttons to move the scanner arm so the laser positioning dot is on the ‘+’ point on the Cal-
ibration Standard Plexiglas surface identified by the inscribed letter “B” and press the MARK button.
Place the QC Phantom in the marked location next to the Calibration Standard, parallel with the backrest.
The 'C' should be to oriented to the right side of the table, (operator facing the table).
Move the scanner arm so that the laser positioning dot is on the dot (or X) by point “C”.
First MARK the dot at point “C”, then MARK the dot at point "D".
The scan of the Calibration Standard proceeds automatically to completion without operator intervention and
the estimated remaining time is updated periodically. Total time is about 25 minutes.
Verify that the PRECISION and ACCURACY fields in the window display the OK status on the BMD, Fat,
and Lean fields.
Click on
to print the results for storing in the calibration log.
Click on Stop and Cancel any time before the QA Results window appears to terminate the calibration pro-
cedure. The calibration files will revert to the previously recorded values.
Note:
Should any part of the calibration process fail, repeat the calibration once. If the diagnostic tests
fail the second time, shut the system down, re-start and perform calibration. If further difficulty is
encountered, contact your Norland Customer Service representative. Refer to the Operator’s Guide
for other troubleshooting options.
Summary of Contents for Elite 437A150
Page 1: ...DXA Bone Densitometer Operator s Guide Elite Model 437A150 Operator s Guide 437D140 Rev I...
Page 22: ...This page intentionally left blank Operator s Guide 437D140 Rev I 1 8...
Page 56: ...This page intentionally left blank Operator s Guide 437D140 Rev I 2 34...
Page 66: ...11 Click Operator s Guide 437D140 Rev I 3 10 Installation and Setup...
Page 108: ...This page intentionally left blank Operator s Guide 437D140 Rev I 3 52...
Page 124: ...Sample Printed QA Results Report Operator s Guide 437D140 Rev I 4 16 Basic Operation...
Page 137: ...Operator s Guide 437D140 Rev I Basic Operation 4 29...
Page 166: ...This page intentionally left blank Operator s Guide 437D140 Rev I 5 24...
Page 198: ...Operator s Guide 437D140 Rev I 7 20 Scanning AP Spine...
Page 206: ...This page intentionally left blank Operator s Guide 437D140 Rev I 7 28...
Page 238: ...Figure 8 6 Page 2 of the Bone Exam Report Operator s Guide 437D140 Rev I 8 32 Scanning Hip...
Page 290: ...This page intentionally left blank Operator s Guide 437D140 Rev I 10 30...
Page 326: ...Operator s Guide 437D140 Rev I 11 36 Additional Techniques...
Page 412: ...This page intentionally left blank Operator s Guide 437D140 Rev I 14 16...
Page 428: ...This page intentionally left blank Operator s Guide 437D140 Rev I 16 10...
Page 446: ...Label Locations continued Operator s Guide 437D140 Rev I 17 18 Technical Reference...
Page 448: ...Scanner Labels continued Operator s Guide 437D140 Rev I 17 20 Technical Reference...
Page 449: ...Operator s Guide 437D140 Rev I Technical Reference 17 21 Scanner Labels continued...
Page 459: ...Operator s Guide 437D140 Rev I Reference Data Sets 18 9...
Page 466: ...Operator s Guide 437D140 Rev I This page intentionally left blank...
Page 484: ...Operator s Guide 437D140 Rev I This page intentionally left blank...