Software Operation
If error messages similar to the image below appear, print the errors or write them down, and contact Technical
Support with the details of the message.
Condition
Action
Fatal Error: Limit
encountered during
command OR Limit
encountered prior to
command
The scanner table has reach the limits at the end of the table before it expected to.
Click OK to close the error window. Turn off the scanner power, and push the scanner
arm toward the middle of the scanner. Turn the scanner power back on. Return to the
Patient Visit List screen. Click Tools > Initialize Scanner to reinitialize the scanner.
“Application already
running” message
when starting Illu-
minatusDXA program
Check in the Windows task bar to see if the IlluminatusDXA software is already run-
ning, and select it to activate the program that is already open. Only one instance of the
program may be running at a time.
If another instance of the program is not currently running, click Yes to restart the Illu-
minatusDXA program.
Time/date display
incorrect
Right-click on the time in the lower right corner of the Windows desktop. Click on
Adjust Date/Time. Adjust as necessary.
Patient scanned under
wrong name, or same
patient has two
records and need to be
merged
Contact Technical Support.
Scan
button not active
Verify that the message “Scanner Present” is displayed at the bottom of the Database
Navigator window in the Norland Illuminatus software. If not, refer to the troubleshoot-
ing condition “No Communication” in the "Startup Faults" on page 1.
When you select
Report and a blank
screen appears
Click Close to close the report screen. Click on Select Template to select the appro-
priate report template. Click Report to generate a new report.
Displays “Scanner Sub-
system is busy” when
trying to exit Illu-
minatus.
Return to the Patient List before exiting.
Table 16-3:
Software Faults
Operator's Guide
437D140 Rev. I
16-4
Troubleshooting
Summary of Contents for Elite 437A150
Page 1: ...DXA Bone Densitometer Operator s Guide Elite Model 437A150 Operator s Guide 437D140 Rev I...
Page 22: ...This page intentionally left blank Operator s Guide 437D140 Rev I 1 8...
Page 56: ...This page intentionally left blank Operator s Guide 437D140 Rev I 2 34...
Page 66: ...11 Click Operator s Guide 437D140 Rev I 3 10 Installation and Setup...
Page 108: ...This page intentionally left blank Operator s Guide 437D140 Rev I 3 52...
Page 124: ...Sample Printed QA Results Report Operator s Guide 437D140 Rev I 4 16 Basic Operation...
Page 137: ...Operator s Guide 437D140 Rev I Basic Operation 4 29...
Page 166: ...This page intentionally left blank Operator s Guide 437D140 Rev I 5 24...
Page 198: ...Operator s Guide 437D140 Rev I 7 20 Scanning AP Spine...
Page 206: ...This page intentionally left blank Operator s Guide 437D140 Rev I 7 28...
Page 238: ...Figure 8 6 Page 2 of the Bone Exam Report Operator s Guide 437D140 Rev I 8 32 Scanning Hip...
Page 290: ...This page intentionally left blank Operator s Guide 437D140 Rev I 10 30...
Page 326: ...Operator s Guide 437D140 Rev I 11 36 Additional Techniques...
Page 412: ...This page intentionally left blank Operator s Guide 437D140 Rev I 14 16...
Page 428: ...This page intentionally left blank Operator s Guide 437D140 Rev I 16 10...
Page 446: ...Label Locations continued Operator s Guide 437D140 Rev I 17 18 Technical Reference...
Page 448: ...Scanner Labels continued Operator s Guide 437D140 Rev I 17 20 Technical Reference...
Page 449: ...Operator s Guide 437D140 Rev I Technical Reference 17 21 Scanner Labels continued...
Page 459: ...Operator s Guide 437D140 Rev I Reference Data Sets 18 9...
Page 466: ...Operator s Guide 437D140 Rev I This page intentionally left blank...
Page 484: ...Operator s Guide 437D140 Rev I This page intentionally left blank...