4-28
Power-Up
and
Testing
7.
In the Reticles Setup dialog box, enter the following sample settings:
•
Circle 1 (0.050)
•
Crosshair graduations (0.025)
•
Circle 2 (0.10)
•
Red reticle color
•
Rectangle (0.10 x 0.10)
8.
Click on the
Apply changes
button
and view the display.
!
If any of the reticles are not visible on the screen, contact Asymtek Technical Support.
9.
Turn off all of the reticle selections and click on the
Apply changes
button.
10.
View the image in the video display.
!
The reticles should not be visible.
!
If the image is focused, the camera does not need to be adjusted.
!
If the image is not focused, the camera focus must be adjusted. See
11.
Verify the bottom of the lighting module is approximately 1.1 inches (27.9 mm) above the
workpiece surface to be viewed with a metal ruler.
!
If the lighting is not at the correct height, grasp the light with your hand and loosen the
adjustment screw. Then slide the light up or down on its bracket until it is at the proper
distance and retighten the adjustment screw. Refer to
12.
Calibrate the camera. See
4.10.8 Dispensing Valve Pneumatics
Perform the following procedure to check function of the dispensing valve pneumatic system.
"
NOTE
The following procedure assumes the dispensing system is on and FmXP is running.
1.
If a dispensing valve is currently installed on your dispensing system, disconnect all air hoses
connected to the dispensing valve.
2.
Connect a syringe receiver head to the silver-colored fluid pressure fitting on the pneumatic
bulkhead located on the right-hand side of the dispensing head.
3.
In the FmXP Main Window, click on
Tools
.
4.
In the Tools Window, click on
I/O Test
and then on
Dispenser
.
!
The Dispenser I/O Test Window opens (Figure 4-16).
5.
Locate the bit labeled
V1_FLUID_AIR
.
6.
Toggle the bit
ON
(
1
).
!
This should turn ON air at the Fluid Pressure port. You should hear air coming out of the
syringe receiver head.
Summary of Contents for Asymtek Spectrum S-920
Page 2: ......
Page 4: ......
Page 30: ...1 14 Introduction 1 10 5 Rear View Figure 1 7A Rear View 2 3 4 1 5...
Page 124: ......
Page 185: ...Troubleshooting 8 11 Figure 8 3 FmXP Scale Setup Menu Figure 8 4 Scale Error Message...
Page 190: ......
Page 204: ......
Page 208: ......
Page 210: ......
Page 220: ......
Page 222: ......
Page 223: ......
Page 224: ......
Page 225: ......
Page 226: ......
Page 234: ......