How to Enable / Disable Low Air Pressure Alarm
Step 1 Press MODE
button and scroll to STEADY.
Step 2 Press and hold
until Aon
or Aof
is visible.
Step 3 Press
or
button to toggle between Alarm On
or Alarm OFF
.
Step 4 Press MODE
button to exit.
How to Select PSI or BAR Pressure Readout
Step 1 Press MODE
button and scroll to STEADY.
Step 2 Press and hold
until Aon
or Aof
is visible.
Step 3 Press
button one time.
Step 4 Press
or
button to toggle between
for BAR, and
for PSI.
PSI Format: 0.0 to 101.0
BAR Format: 0.0 to 7.0
Step 5 Press MODE button to exit.
How to Enable / Disable CC INIT I/O as an External Alarm Input
Step 1 Press MODE button and scroll to STEADY.
Step 2 Press and hold
until Aon
or Aof
is visible.
Step 3 Press
two times.
Step 4 Press
or
button to toggle between CCI
or ALI
CC INIT function is:
CCI:
Contact Closure initiate input
ALI:
External alarm input
Step 5 Press MODE button to exit.