Ultimus I-III Dispenser Pressure Readout, 
Vacuum Readout, and Dispense Timer Validation

Instructions

Introduction

These instructions provide the pressure, vacuum, and dispense timer validation procedures for the following dispensers:
• Ultimus

 I: P/N 7017041  

• Ultimus II: P/N 7002003   

• Ultimus III P/N 7017068 (legacy)

Specification List

Model

Pressure Accuracy

Vacuum Accuracy

Timer Accuracy

Ultimus I

±2.0 psi, 0–100 psi

±2.0 inH

2

O, 0–6 inH

2

O

±0.05%

Ultimus II

±0.3 psi, 0–15 psi

±2.0 inH

2

O, 0–6 inH

2

O

±0.05%

Ultimus III

±0.3 psi, 0–5 psi

±2.0 inH

2

O, 0–6 inH

2

O

±0.05%

Required Tools and Supplies

The following equipment is required for the performance of these procedures.

#

Description

Minimum Accuracy Requirement

Suggested 

Equipment

Purpose

1*

Precision pressure 
and vacuum test 
gauge

±0.05 psi at 0–100 psi 
±0.33" H

2

O

Fluke, model 700G06 
pressure and vacuum 
test gauge

Reference standard for 100/15 psi 
pressure and 6" H

2

O vacuum

2

0–100 psi test gauge, 
±0.25% full scale 
accuracy

±0.25 psi

WIKA

®

, model 312.20 6, 

P/N 9746892

Reference standard for validating the 
Ultimus I pressure readout

3

0–15 psi test gauge, 
±0.25% full scale 
accuracy

±0.05 psi

WIKA

®

, model 312.20 6, 

P/N 9746867

Reference standard for validating the 
Ultimus II and III pressure readouts

4

Slack tube 
manometer, 18-0-18"

±0.33" H

2

O

Dwyer Instruments, 
model 1211-36

Reference standard for validating the 
Ultimus I, II, and III vacuum readouts

5

Function generator, 
triggered 10 s pulse, 
±1 ppm accuracy

±20 ppm for a 10 s triggered pulse; 
5–24V for ON, 
0V for OFF 
Load: 1,200 ohms

Tektronix

®

, AFG 3021B

Reference standard for validating the 
Ultimus I, II, and III dispense timer

*Item 1 is the preferred equipment and can be used in place of items 2-4. When item 1 is used, only item 5 is also required.

Optional Accessories

The following recommended Nordson EFD accessories are available to support the performance validation and field validation of the 
vacuum display.

Accessory

Description

Nordson EFD 

Part Number

Purpose

Input / output (I/O) cable, 
Ultimus I, II, III

8-pin DIN cable, color coded 
leads, stripped and tinned lead 
ends, 1.9 m (6 ft)

7017143

Connection to Ultimus I, II, or III from the function 
generator for validation of the dispense timer

Vacuum calibration kit, 
Ultimus I, II, III 

Nordson EFD vacuum 
calibration software, USB to 
RS-232 converter, RS-232 to 
modular plug cable

7028817

Adjustment of the Ultimus I, II, or III vacuum 
readout validation; kit provides hardware and 
software for site-specific validation of the digital 
vacuum readout