EV Series Automated Dispensing Systems
37
Setting Up and Calibrating the System (Required)
Before creating any programs or using the automatic offset update capabilities of the system, you must properly set
up and calibrate the system. Correct system setup and calibration is critical for proper system operation and must
be performed as follows:
• At initial system startup.
• After a component on the Z axis (such as the syringe barrel or camera) is moved or if the relationship between
the dispensing tip and/or camera is altered.
Setup and calibration includes the following procedures:
• Verifying the system configuration.
• Focusing/calibrating the camera.
• Calibrating and setting the camera-to-tip offset.
• Calibrating and setting the tip-to-workpiece offset.
• Testing the setup.
NOTE:
Refer to “About Offsets” on page 23 for an explanation of offsets.
Verifying the Robot Model and Tip Detector Selection
#
Click
Step
Reference Image
1
>
• Click SYSTEM SETUP > OPEN.
2
• Under Other, verify the following:
- The correct robot model is selected.
- If your system includes a tip detector, Tip
Detect Device is checked.
• If you made changes, close and reopen the
DispenseMotion software for the changes to
take effect.