75
II. Alarm message
“communication with
the oven was lost”
Check
For:
A. Power
For HC1 Control board:
1. Verify
proper
operation
of “Secondary Power” (
See
Troubleshooting Guide
,
section I.B
)
2. Check the power input between the terminals marked J1-1 and
J1-2. It should be approximately 230VAC (
See
Figure 4a
).
3. Make sure fuses F1 & F2 (0.25A) are not open (
See
Figure 4a
).
4. Check the proper connection of communication cable and also
check wiring for communication cable. Communication cable
should be connected at J13.
For HC2 Control board:
1. Check the power input between the terminals marked J1+ and
J1-. It should be approximately 24VDC (
See
Figure 4b
).
2. Make sure fuse F1 (1A) is not open (
See
Figure 4b
).
3. Check the communication cable, it should be connected at J1 of
the HDB board which is Ethernet port to the computer Ethernet
port.