Nonin SenSmart X-100 Service Training Download Page 1

Summary of Contents for SenSmart X-100

Page 1: ......

Page 2: ...Signal Processor 7 System Sensors 7 Connect Disconnect a Sensor to from the Signal Processor 7 rSO2 Sensors 7 rSO2 Sensor Application 8 SpO2 Sensors 9 Reusable Pulse Oximetry SpO2 Sensor Application 9...

Page 3: ...se Menu 21 System Menu 22 Review System Menus 24 System Alarm Indicators 25 High Priority Alarms 25 Medium Priority Alarms 25 Troubleshooting 26 System Memory 28 System Data Outputs 28 Cleaning the Sy...

Page 4: ...stem components and functions 2 Learn what actions are needed to make the system function properly and 3 Provide you the opportunity to demonstrate what you have learned Training format conventions in...

Page 5: ...oximetry SpO2 functionality is suitable for use in both motion and non motion conditions including patients who are well or poorly perfused WARNING This device is intended only as an adjunct device i...

Page 6: ...oxygen saturation changes thus providing early warning of dangerous hypoxemia A pulse oximeter shines light composed of two wavelengths red and infrared through a part of the body that is relatively...

Page 7: ...to bed rails and equipment poles D SenSmart X 100SP X Oximetry Signal Processors 1 through 6 Each Signal Processor is programmed to be a specific channel on the Monitor designated by number and can b...

Page 8: ...sed to eliminate the potential for environmental contamination Connect Disconnect the Hub to the Monitor 1 To connect a Align the arrow on the Hub cable connector with the small triangle on the Monito...

Page 9: ...ensor lock to secure the sensor firmly in the Signal Processor See Figure 5 The measurement from the Signal Processor will display on the Monitor screen in the same color and channel location as shown...

Page 10: ...To connect a Flip the clear sensor lock on the Signal Processor back to expose the sensor port b Align the white arrow on the sensor connector with the white arrow on the Signal Processor sensor port...

Page 11: ...t and neonate pediatric patients b The area s should be free of hair or surface blemishes such as moles or freckles c Avoid placing the sensor s over nevi sinus cavities hematomas or arteriovenous mal...

Page 12: ...wound incision compromised skin or pre existing skin condition e g eczema or dermatitis CAUTION Do not use if patient has pre existing sensitivity to adhesive tapes or adhesives on electrocardiogram...

Page 13: ...nector port as shown in Figure 11 b Push the Extension Cable connector into the Hub port or Monitor connector port until it clicks and locks into place c Connect the other end of the Extension Cable i...

Page 14: ...pole using the clip on the back of the Hub Holster 4 Connect three 3 Signal Processors to the Hub 5 Connect two 2 rSO2 sensors to the Signal Processor and lock into place 6 Connect one 1 SpO2 sensor t...

Page 15: ...for each channel with rSO2 trend lines or SpO2 plethysmograms C Channel values rSO2 or SpO2 data for each active channel D Function control buttons Function buttons to control specific system functio...

Page 16: ...lect item to edit b Save entered or changed values B Alarm Silence 1 Press once to silence audible alarms for 2 minutes 2 Press once to reactivate audible alarms before the 2 minute silence period is...

Page 17: ...E Baseline 1 Press twice to quickly set all baseline s to the current rSO2 values 2 rSO2 Baseline values can be set by the user for each patient to compare subsequent rSO2 changes relative to this Ba...

Page 18: ...E AUC area under the curve Society for Thoracic Surgery STS database reporting value automatically calculated as minutes multiplied by the percent below the set lower alarm limit See page 16 Figure 16...

Page 19: ...e which shows the last 10 event marks date time and values for each active channel See Figure 17 below Scrolling Cursor and Values R Cursor yellow down arrow and chart of channel values Cursor yellow...

Page 20: ...Type Select the type of sensor for each channel rSO2 or SpO2 Baseline rSO2 only Set Baseline manually numerical value rSO2 Baseline values can be set by the user for each patient to compare subsequent...

Page 21: ...omatically defined and disabled gray when a compatible sensor is attached to the Signal Processor Baseline Manual Setting rSO2 Baseline values can be set by the user for each patient to compare subseq...

Page 22: ...situations Do not cover or otherwise block the speaker opening Graph Position Data can be graphed for each channel on the display Regional oximetry values will be graphed as a trend line pulse oximet...

Page 23: ...Preset and pressing Menu again returns to the monitoring screen Presets Menu A Preset is a collection of channel specific user defined or default settings configured in the Settings menu and the firs...

Page 24: ...o move to and highlight Start new case 2 Press Select Start new case pop up displays with No highlighted 3 Press the down navigation arrow to highlight Yes 4 Press Select If system is set up to enter...

Page 25: ...recordings from Monitor and attached signal processors Restore Factory Defaults Discards all Presets and institution settings and restores factory defaults System Information Hardware and software ve...

Page 26: ...settings See the SenSmart System Operator s Manual for complete instructions for use Setting the Date and Time 1 Select the System menu 2 Select the Date Time option in the second column 3 Select the...

Page 27: ...a rSO2 Low alarm setting of 25 j Set both rSO2 channels to a rSO2 High alarm setting of 90 2 Verify your new settings in the Settings Menu 3 Save these new settings as a Preset named TRAINING in the...

Page 28: ...5 second pause cycle repeats until silenced WARNING Ensure all alarm volumes are set appropriately and are audible in all situations Do not cover or otherwise block any speaker openings Medium Priorit...

Page 29: ...Sensor or Signal Processor is disconnected 1 Ensure all connections between components are secure a Check the connection between the sensor and the Signal Processor b Check the connection between the...

Page 30: ...Nonin Technical Service ONE OR MORE CHANNELS DISPLAY MESSAGE UNRECOGNIZED PACKET One or more channels display the message Unrecognized packet Verify that duplicate signal processors are not attached...

Page 31: ...en the memory capacity is reached as described in Table 13 below the System begins overwriting the oldest data with new data rSO2 or SpO2 and HR is sampled and recorded once every 4 seconds for each c...

Page 32: ...ach solution household bleach 5 25 sodium hypochlorite Do not use undiluted bleach or any cleaning solution other than those recommended here as permanent damage could result 2 Dry with a soft cloth o...

Page 33: ...to Get Help Service and Support Nonin Medical offers extensive technical clinical and customer service support for questions or assistance These include Your local Account Manager Clinical Specialist...

Page 34: ...SenSmart Model X 100 Universal Oximetry System In Service Training 31 of 31...