Data Output and Software
Data Output and Software
Device Memory
The monitor can collect and store a minimum of 36 hours of patient data, including 8 hours of
waveform. To store data in memory, the monitoring duration must be at least 2 minutes.
Data may be downloaded from the monitor using a USB flash drive. Memory may be cleared once
data is downloaded.
Patient data are sampled every second.
•
EtCO
2
values are stored in 1 mmHg (0.1 kPa) increments in the range of 0 – 99 mmHg
(0 – 13.2 kPa).
•
Respiration rate values are stored in 1 BrPM increments in the range of 0 – 99 BrPM.
Patient data are retained even when both external and battery power are lost.
The memory in the device functions much like an “endless loop” tape. When the memory is full,
the monitor begins overwriting the oldest data with new data.
Download Patient Data
USB Drive Requirements
•
Formatted with FAT32 file system
•
Compatible with USB 2.0 or 2.1
•
More than 1 GB available space
•
100 mA maximum current draw
Recommended SanDisk
®
USB Flash Drives
Nonin has tested these flash drives with the monitor:
NOTE:
If USB flash drive performance issues occur, see the
section for
more information before contacting Nonin. Nonin cannot guarantee USB flash drive
performance if a recommended flash drive is not used.
CAUTION:
If the entire memory is filled, portions of the oldest record will be
overwritten when new data is written.
Name
Size
SanDisk P/N
Cruzer Glide™
8 GB
section)
Cruzer Edge™
8 GB
SDCZ51-008G-A11
Cruzer Force™
8 GB
SDCZ71-008G-A46
Cruzer Blade™
4 GB
SDCZ50-004G-A46
!