23
Using the RespSense II Monitor
10. When done monitoring, press
On/Standby
for 3 seconds to turn off the monitor.
11. Disconnect the patient.
NOTE:
If the monitor is on and the patient is no longer connected, the alarm will activate.
Adjusting Display Brightness
The display brightness can be adjusted while monitoring. The default brightness is maximum.
1. In the lower left corner of the operating screen, press the brightness icon
.
2. With each press of the icon, the screen brightness changes from maximum to medium to
low.
Adjusting Audible Alarm Volume
The alarm volume can be adjusted while monitoring. The default volume is maximum.
1. In the lower right corner of the operating screen, press the audible alarm volume icon
.
2. With each press of the icon, the audible alarm volume changes from off to medium to
maximum (if allowed by the Responsible Organization Settings).
NOTES:
•
Configure the alarm volume options using the Responsible Organization Settings screen
(see the
sections).
•
Audible alarms cannot be shut off if the Minimum Alarm Volume field is set to Medium or
Maximum.
•
The default alarm volume is either the saved Default Alarm Volume or the last used
volume (if Last Used Alarm Setting is selected at start-up).
CAUTION:
Set or adjust only one parameter at a time.
WARNING: The device is intended only as an adjunct device in patient assessment. It
must be used in conjunction with other methods of assessing clinical signs and
symptoms.
!