Operator’s Manual
Avant
®
4000 Digital
Pulse Oximetry System
with Bluetooth
Wireless Technology
Avant 4000
0123
English
Page 1: ...Operator s Manual Avant 4000 Digital Pulse Oximetry System with Bluetooth Wireless Technology Avant 4000 0123 English...
Page 2: ...ithout notice or obligation Nonin Medical Inc 2605 Fernbrook Lane North Plymouth Minnesota 55447 4755 USA 1 763 553 9968 800 356 8874 USA and Canada Fax 1 763 553 7807 mail nonin com www nonin com Aut...
Page 3: ...14 Verifying System Operation 16 Device Pairing 17 Default Settings 18 Accessing User Functions 19 Display Unit DIP switches 22 Care and Maintenance 23 Alarms and Limits 24 Alarms and Informational To...
Page 4: ...materials Type BF Applied Part Patient isolation from electrical shock UL Mark for Canada and the United States with respect to electric shock fire and mechanical hazards only in accordance with UL 2...
Page 5: ...angulation All parts and accessories connected to the serial port of this system must be certified according to IEC Standard EN 60950 or UL 1950 for data processing equipment To avoid the risk of conf...
Page 6: ...te at least every 6 to 8 hours to ensure correct sensor alignment and skin integrity Patient sensitivity to sensors may vary due to medical status or skin condition Do not place liquids on top of this...
Page 7: ...g measures 1 Reorient or relocate the receiving antenna 2 Increase the distance between the equipment and the receiver 3 Connect the equipment to an outlet on a circuit different from the outlet where...
Page 8: ...es 8000AA WO finger clip sensor Operator s manual Universal Desktop Battery Charger with IEC320 Connector 3 wrist bands Indications for Use The NONIN Avant 4000 Digital Pulse Oximetry System is indica...
Page 9: ...ate and plethysmographic data to be transmitted through a Bluetooth radio to a compatible Bluetooth enabled device Nonin s system removes the connection from the sensor cable to the display unit givin...
Page 10: ...necessary to operate the device without batteries audible alarms and memory functions may not be available DO NOT use the device without batteries when patient safety relies upon audible alarms NOTE...
Page 11: ...8 Using the Avant 4000 Digital Pulse Oximetry System Installing Batteries in the Patient Module...
Page 12: ...ED 7 Numeric LEDs for Pulse Rate data 8 Alarm Silence Button 9 Alarm Limits Button and Indicator 10 Plus Button 11 Minus Button 12 Time LED 13 Month LED 14 Day LED 15 Year LED 16 Pulse Volume LED 17 A...
Page 13: ...s these LEDs display in green For high priority patient alarms the corresponding values are displayed in red blinking fast The values are displayed in amber for medium priority alarms and when reviewi...
Page 14: ...lect from multiple wrist worn patient modules during pairing Memory Button This button is used to enter Memory Playback mode Bluetooth Control Button Press and hold this button while turning on the di...
Page 15: ...by the oximeter The height of the Pulse Strength Bargraph LED is proportional to the pulse signal and the color is determined by pulse strength Green a good pulse strength Amber a marginal pulse stren...
Page 16: ...l it remains lit steadily AC Power Adapter LED This green LED is displayed when an external power supply is providing power to the display unit Battery LED This amber LED indicates a marginal battery...
Page 17: ...in the display unit and wrist worn patient module WARNING The battery pack must be installed at all times while the device is operating even when operating on AC power If it is necessary to operate th...
Page 18: ...n of the display unit and pair the display unit with the desired wrist worn patient module See Verifying System Operation and Device Pairing for more information 6 Secure the wristband to the patient...
Page 19: ...he next page The Main Alarm LED and Pulse Oximeter LED blink amber until the oximeter produces valid readings Use the procedure below to monitor SpO2 and pulse rate readings in order to verify that th...
Page 20: ...mum of five pairable patient modules will be displayed separately with the device identification number appearing in the SpO2 and Pulse Rate display areas A unique device identification number is incl...
Page 21: ...olume adjustments made while the device is operating in Factory Default Setting will be lost when the device is shut down The last adjusted values can be recalled by pressing and holding both the Alar...
Page 22: ...Check the battery charge Press the ON STANDBY button while the unit is on Battery charge is displayed in 10 increments for 4 seconds in green on the Pulse Strength Bargraph Mute the audible alarms 2...
Page 23: ...s Press the Alarm Limits button to step through the Limits menu Use the Plus or Minus buttons to adjust alarm limits Set pulse and alarm volumes Press the Volume button to select pulse or alarm volume...
Page 24: ...with the device identification number appearing in the pulse rate and SpO2 display areas When the desired patient module s identification number is displayed pressing the Bluetooth Control button com...
Page 25: ...me cannot be disabled Switch 3 Date Format Up International Date format Day Month Year Down U S A Date format Month Day Year Switch 4 Data Output Up Real time once per second data output Down Print on...
Page 26: ...bleshooting Cleaning the Avant 4000 Digital Pulse Oximetry System Clean all system components with a soft cloth dampened with isopropyl alcohol Do not pour or spray any liquids onto components and do...
Page 27: ...alarms are sounded as follows beep beep beep short pause beep beep 10 second pause Medium Priority Alarms Medium priority alarms are those that signal potential problems with the equipment or other no...
Page 28: ...rder to permanently silence all alarms DIP switch 2 must be placed in the UP position This allows the alarm volume to be set to zero The Alarm Silence LED will remain illuminated when the alarm volume...
Page 29: ...lly after ten seconds with no activity 3 Ensure that the appropriate upper or lower Alarm Limit LED is illuminated and that the alarm limit you want to change is displayed 4 Press the Plus or Minus bu...
Page 30: ...steps 1 Turn the unit off and then back on again to remove the error code 2 If the error persists disconnect all power AC and battery and then reconnect the power and turn the unit back on 3 If the er...
Page 31: ...is stored in memory starting a new recording session See Displays Indicators and Controls for more information about setting the clock Only recording sessions greater than one minute in length are st...
Page 32: ...se Rate LED display areas This message signals that the device is in Playback mode 3 The PLy bAC message will disappear when memory playback is complete Memory playback may take up to 8 minutes depend...
Page 33: ...em in one of the following formats If DIP switch 8 is in the DOWN position the data will be displayed as follows SPO2 XXX HR YYY where XXX is the SpO2 value and YYY is the heart rate value If DIP swit...
Page 34: ...real time once per second data output is available via the RS232 connector port using a null modem cable When DIP switch 4 is in the DOWN position default the Print on Demand mode is activated allowi...
Page 35: ...anometers 3 mw nominal Internal Power Battery Two 1 5 volt AA batteries Operating Life minimum 120 hours of continuous operation with new batteries Storage Life 10 months Weight 4 4 ounces with batter...
Page 36: ...w at least one hour of stabilization time before use Antenna Type Inverted F type antenna Antenna Gain 2 dB typ 3 dB max TRANSMITTER Bluetooth Compliance Version 1 1 Operating Frequency 2 4 to 2 4835...
Page 37: ...nector Contact your distributor or NONIN for options Cord Set Charger External Cables UNI RS232 RS 232 Cable Pulse Oximeter Sensors 8000AA WO Adult Articulated Finger Clip Sensor 8000J WO Adult Flex P...
Page 38: ...sive remedy by the purchaser hereunder for any Avant 4000 Digital Pulse Oximetry System delivered to the purchaser which is found to be defective in any manner whether such remedies be in contract tor...
Page 39: ...Reposition the digit or insert a different digit and keep the sensor motionless for at least 10 seconds Warm the digit by rubbing or covering with a blanket Position the sensor at a different site NOT...
Page 40: ...to the 4100 Check the 4100 s batteries Check the sensor for any visible signs of deterioration Contact your distributor or NONIN Customer Support Frequent or steady pulse quality indication There is e...
Page 41: ...connections and retry Verify that the 4100 is paired with the display Verify that the Connection Status LED is green on both the display unit and the patient module Contact your distributor or NONIN...
Page 42: ...IP switch 2 to the DOWN position if you desire audible alarms The system is not functioning correctly Contact your distributor or NONIN Customer Support The devices will not pair The sensor has not be...