Fast wavelength modulation
38
Table 6
Wavelength modulation settings
Wav/-
signal
The
Wavelength
) are a differential input/output used for
wavelength modulation of the laser. You can configure the wavelength
modulation option so that the interface is used either as input or output. It is
preferential to use a differential signal for the modulating signal applied. When
using a differential signal, the combined voltage of the
Wavelength
+ and
Wavelength
- pins modifies the wavelength with the following results:
•
a negative voltage results in a lower output wavelength.
•
whereas a positive voltage results in a higher output wavelength.
Setting
Function
For CONTROL see For SDK see
External wavelength
modulation
Sets wavelength
modulation to external
signal mode.
Internal function
generator
Sets wavelength
modulation to internal
function generator mode.
Modulation
frequency
Sets the frequency of the
internal function
generator. The settable
range is from 0.008 Hz to
100 kHz.
Modulation
amplitude
Sets the modulation
maximum amplitude to
between 0 & 1000% when
using the internal function
generator mode.
Waveform
When set to internal
function generator mode,
waveform sets the signal
type to either sinusoidal or
sawtooth.
Signal output
Sets the wav/-
pins to output the internal
function generator signal.
Wavelength
modulation
Toggles all wavelength
modulation functionality
and settings on or off.
2
Summary of Contents for Koheras ADJUSTIK HP
Page 1: ...Item 800 611 01 Koheras ADJUSTIK HP PRODUCT GUIDE Low Noise Single Frequency Laser System...
Page 10: ...10 E Part numbers 113...
Page 12: ...12...
Page 16: ...16...
Page 18: ...18...
Page 20: ...20...
Page 36: ...Chassis labels 36...
Page 48: ...48...
Page 56: ...Menu items 56...
Page 68: ...Enabling and disabling emission 68...
Page 84: ...CONTROL Control panel 84...
Page 86: ...86...
Page 90: ...Placing the laser on a table or shelf 90...
Page 98: ...98...
Page 102: ...102...
Page 112: ...Installing CONTROL 112...
Page 114: ...114 2...
Page 115: ...Koheras ADJUSTIK Product Description Release 1 0 04 2021 W 10456...