Heat dissipation
30
Table 3
BOOSTIK installation recommendations
BOOSTIK operating temperature
In general, the actual operating temperature of a BOOSTIK module inserted in an
ACOUSTIK system depends on the following:
•
Ambient temperature
•
Overall airflow around the ACOUSTIK.
•
Total number of inserted and operational modules.
•
Connected seed laser module type: C15. E15, X15, or Y10
•
Actual output power level of the BOOSTIK.
•
Slot spacing between modules.
Caution:
Only insert the amplifier modules in slots according to the configura-
. Operating the ACOUSTIK with more than two
BOOSTIK modules inserted, or inserting BOOSTIK modules in slots not matching
the configurations in
, may result in thermal damage to the ACOUSTIK or
its inserted modules.
Figure 8
Rear panel heat sink
Item
Recommendation
Value
# of BOOSTIK modules
Maximum 2 BOOSTIK modules
per ACOUSTIK system.
≤ 2
Installation slots
.
Do not install two BOOSTIK
modules in adjacent slots.
Slots 15-16 not recommended.
Slots
1-5
6-10
11-14
Operating temperature
<
50°C
Ambient temperature of the
installation environment
Airflow type
R
estricted air flow
Free air flow
Fan-forced air flow
≤
30°C
≤
35°C
> 35°C
Maintain good airflow
around the heat sink cooling fins
Summary of Contents for Koheras ACOUSTIK
Page 1: ...Item 800 633 01 Koheras ACOUSTIK PRODUCT GUIDE Multi Channel Modular Laser System...
Page 10: ...10...
Page 14: ...14...
Page 16: ...16...
Page 26: ...System labels 26...
Page 28: ...28...
Page 34: ...Operating and storage environment 34...
Page 42: ...42...
Page 78: ...Module overview 78...
Page 80: ...80...
Page 82: ...82 Figure 44 Mechanical dimensions 483 00 45 00 302 50 4 50 125 00 2...
Page 92: ...Installing CONTROL 92...
Page 95: ......
Page 96: ...1 Koheras ACOUSTIK Product Guide Revision 1 0 09 2021 W 10456...