Instruction Manual
NivuFlow Mobile 750
page 30
NFM 750 - rev. 03 / 27.05.2020
21.2.2
Water ultrasound
The Type CSP-V2H1 and CSP-V2U1 sensors are equipped with a horizontal crystal element
used for level measurement based on the transit time method using water-ultrasound.
The measurement uses the time between transmission and reception of an impulse being re-
flected on the water surface.
ℎ
1
=
1
2 ×
(
𝑐𝑐
×
𝑡𝑡
1
)
•
with:
h = filling level
c = sound running time
t
1
= time between transmitted and received signal
Fig. 21-6
Determined filling level
At a medium temperature of 20 °C the sound running time in water is 1480 m/s.
The temperature-dependent deviation is 0.23 % per Kelvin.
To achieve level measurements with millimetre accuracy, the medium temperature is investi-
gated permanently. The investigated temperature is used to correct the sound running time
for calculation purposes.
The fixed level value is added to the determined value h
1
. The result is the total flow level h.
21.2.3
Air-ultrasound
Level is measured via air-ultrasound from top down by using a contactless air-ultrasonic sen-
sor Type DSM-L0. This sensor is used to measure low filling levels. It must be installed exact-
ly in the centre of the channel crown (±2°) parallel to the water surface.
Analogous to the transit time method based on water-ultrasound, a signal is transmitted too.
In this case, however, the signal is sent through the air towards the water surface from where
it is reflected. The measured transit time provides the distance between sensor and water
surface and hence the filling level of the medium.
Since the air temperature may lead to deviations in the transit time the temperature is deter-
mined and will be considered during internal calculations.
Summary of Contents for NivuFlow 750
Page 141: ...Approvals and Certificates NFM 750 rev 03 27 05 2020 page 141 Approvals and Certificates...
Page 142: ...Instruction Manual NivuFlow Mobile 750 page 142 NFM 750 rev 03 27 05 2020...
Page 143: ...Approvals and Certificates NFM 750 rev 03 27 05 2020 page 143...
Page 144: ...Instruction Manual NivuFlow Mobile 750 page 144 NFM 750 rev 03 27 05 2020...
Page 145: ...Approvals and Certificates NFM 750 rev 03 27 05 2020 page 145...
Page 146: ...Instruction Manual NivuFlow Mobile 750 page 146 NFM 750 rev 03 27 05 2020...
Page 147: ...Approvals and Certificates NFM 750 rev 03 27 05 2020 page 147...
Page 148: ...Instruction Manual NivuFlow Mobile 750 page 148 NFM 750 rev 03 27 05 2020...
Page 149: ...Approvals and Certificates NFM 750 rev 03 27 05 2020 page 149...
Page 150: ...Instruction Manual NivuFlow Mobile 750 page 150 NFM 750 rev 03 27 05 2020...
Page 151: ...Approvals and Certificates NFM 750 rev 03 27 05 2020 page 151...
Page 152: ...Instruction Manual NivuFlow Mobile 750 page 152 NFM 750 rev 03 27 05 2020...
Page 153: ...Approvals and Certificates NFM 750 rev 03 27 05 2020 page 153...
Page 154: ...Instruction Manual NivuFlow Mobile 750 page 154 NFM 750 rev 03 27 05 2020...
Page 155: ...Approvals and Certificates NFM 750 rev 03 27 05 2020 page 155...
Page 156: ...Instruction Manual NivuFlow Mobile 750 page 156 NFM 750 rev 03 27 05 2020...
Page 157: ...Approvals and Certificates NFM 750 rev 03 27 05 2020 page 157...
Page 158: ...Instruction Manual NivuFlow Mobile 750 page 158 NFM 750 rev 03 27 05 2020...
Page 159: ...Approvals and Certificates NFM 750 rev 03 27 05 2020 page 159...
Page 160: ...Instruction Manual NivuFlow Mobile 750 page 160 NFM 750 rev 03 27 05 2020...
Page 161: ...Approvals and Certificates NFM 750 rev 03 27 05 2020 page 161...
Page 162: ...Instruction Manual NivuFlow Mobile 750 page 162 NFM 750 rev 03 27 05 2020...