Instruction Manual
NivuFlow Mobile 750
page 18
NFM 750 - rev. 03 / 27.05.2020
8
Ex Protection
The portable transmitter NivuFlow Mobile 750 and the accompanying sensors are designed
to be used in areas with explosive atmospheres (zone 1).
The following conditions must be observed:
•
Programming the unit under Ex conditions is permissible:
with the programming person
within
the Ex area as soon as the indicator and
operating unit used has an Ex approval
with the programming person
outside
of the Ex area as soon as the indicator
and operating unit used has no Ex approval
•
Maintenance and repairs shall be executed only
outside
of Ex area
•
Rechargeable batteries shall be removed/inserted and charged only
outside
of
Ex areas
•
In general, only rechargeable batteries approved by NIVUS shall be used
within
Ex areas
•
Use the USB interface only
outside
of Ex areas
•
Replace the SIM card only
outside
of Ex areas
•
The instrument shall be (after successful installation at the measurement place) se-
cured against unauthorised opening by using a padlock (holes on the side of the enclo-
sure)
Approval
Transmitter
II 2G Ex eb ib [ib] mb IIB T4 Gb (TÜV 17 ATEX 196722 X)
Sensors
II 2G Ex ib IIB T4 Gb (TÜV 12 ATEX 087812)
Validity of Ex Approval
The Ex approval is only valid in connection with the respective indication on the transmitter
or the sensor nameplate.
Declarations of Conformity and Test Reports
For installation and commissioning the EC Declarations of Conformity and Test Reports of
the respective authorities shall be strictly followed.
Ex Approval for Sensors
The Ex approval for the sensors is part of the Technical Description for Cross Correlation
and Doppler Sensors.
Summary of Contents for NivuFlow 750
Page 141: ...Approvals and Certificates NFM 750 rev 03 27 05 2020 page 141 Approvals and Certificates...
Page 142: ...Instruction Manual NivuFlow Mobile 750 page 142 NFM 750 rev 03 27 05 2020...
Page 143: ...Approvals and Certificates NFM 750 rev 03 27 05 2020 page 143...
Page 144: ...Instruction Manual NivuFlow Mobile 750 page 144 NFM 750 rev 03 27 05 2020...
Page 145: ...Approvals and Certificates NFM 750 rev 03 27 05 2020 page 145...
Page 146: ...Instruction Manual NivuFlow Mobile 750 page 146 NFM 750 rev 03 27 05 2020...
Page 147: ...Approvals and Certificates NFM 750 rev 03 27 05 2020 page 147...
Page 148: ...Instruction Manual NivuFlow Mobile 750 page 148 NFM 750 rev 03 27 05 2020...
Page 149: ...Approvals and Certificates NFM 750 rev 03 27 05 2020 page 149...
Page 150: ...Instruction Manual NivuFlow Mobile 750 page 150 NFM 750 rev 03 27 05 2020...
Page 151: ...Approvals and Certificates NFM 750 rev 03 27 05 2020 page 151...
Page 152: ...Instruction Manual NivuFlow Mobile 750 page 152 NFM 750 rev 03 27 05 2020...
Page 153: ...Approvals and Certificates NFM 750 rev 03 27 05 2020 page 153...
Page 154: ...Instruction Manual NivuFlow Mobile 750 page 154 NFM 750 rev 03 27 05 2020...
Page 155: ...Approvals and Certificates NFM 750 rev 03 27 05 2020 page 155...
Page 156: ...Instruction Manual NivuFlow Mobile 750 page 156 NFM 750 rev 03 27 05 2020...
Page 157: ...Approvals and Certificates NFM 750 rev 03 27 05 2020 page 157...
Page 158: ...Instruction Manual NivuFlow Mobile 750 page 158 NFM 750 rev 03 27 05 2020...
Page 159: ...Approvals and Certificates NFM 750 rev 03 27 05 2020 page 159...
Page 160: ...Instruction Manual NivuFlow Mobile 750 page 160 NFM 750 rev 03 27 05 2020...
Page 161: ...Approvals and Certificates NFM 750 rev 03 27 05 2020 page 161...
Page 162: ...Instruction Manual NivuFlow Mobile 750 page 162 NFM 750 rev 03 27 05 2020...