Setting Parameters
page 113
NF 650 - rev. 01 / 29.06.2020
Fig. 33-7
Transmission period / data depth / compression
9. To choose the desired data format turn the rotary pushbutton which opens a selec-
tion menu.
The formats txt and csv are available.
10. Press the rotary pushbutton to accept the data format.
The selectable
data depth
comprises five possible levels:
•
Standard
This is the appropriate format for most applications and hence is the default setting.
The saved data sets contain the following information:
Date and time
Totaliser
Calculated flow rate
Level
Average flow velocity
Water temperature
Current values as well as the accordingly calculated values of enabled analog
and digital inputs
•
Extended
This option is appropriate for the verification of critical, important applications and is
required mainly for servicing personnel.
The saved data sets contain the following information:
All data sets from the previous data depth >Standard<
Average flow velocities of the v-Paths
•
Expert
This option is appropriate for the verification of critical, important applications and is
required mainly for servicing personnel.
The saved data sets contain the following information:
All data sets from the previous data depth >Extended<
Noise
Amplification
•
Day Totals
This option saves only the day totals, no individual values.