33.4 USB stick
Requirements to USB sticks:
• USB 2.0 supported
• FAT 32 format (or FAT 12 or FAT 16)
• Maximum permissible memory 32 GB
Working with USB stick:
Â
Ü
Plug the USB stick into the USB slot located above the display.
Functions:
• Transmission of measurement data to USB stick
• Instrument parameters can be saved to USB stick
• Saved parameters can be restored from USB stick back to the instrument
• Formatting the USB stick
Fig. 33-6
Selection submenu
The transmitter is equipped with an internal data memory. Portions of your measurement data
or the complete memory contents can be transmitted to USB stick.
Within this section it is possible to specify the desired period of transmission.
Per default
the transmitter is set to transmit the data between the latest transmission and the
current point in time. You are free to adjust this period as desired, however.
Â
Ü
To
transmit data
to USB stick proceed as follows:
1. Press the rotary pushbutton to engage the top field.
2. Turn the rotary pushbutton to specify the desired start day.
3. Press the rotary pushbutton again to get to the month input.
4. Repeat the procedure until the desired date and time have been specified comple-
tely.
5. Confirm your entries by pressing the right function key >Enter<.
6. Turn the rotary pushbutton to highlight the >to< field blue.
7. Press the rotary pushbutton to set the desired stop day.
8. Specify the desired stop time as described before.
Now the period of data to be transmitted to USB stick is specified.