![NIVELCO NIPRESS DRD-6 Series User Manual Download Page 1](http://html1.mh-extra.com/html/nivelco/nipress-drd-6-series/nipress-drd-6-series_user-manual_1696807001.webp)
drd6322a0600h_01 - 1/3
Thank you for choosing NIVELCO instrument
We are convinced that you will be satisfied with our product!
1. APPLICATION
NIPRESS D-600
series pressure transmitters with ceramic flush sensor measure pressure and convert it into
voltage and current output can be used in either 2- and 3-wire systems. The
D-600 transmitters are suitable for
the pressure measurement of fluids, oils, and gases. Due to their flush sensor, they are ideal for the measurement
of viscose and polluted media. For using in aggressive media, we recommend PVDF process connection.
Transmitters can be delivered with plug in display
UNICONT PLK-501
enabling on site reading. Ex or SIL
versions are also available.
2. TECHNICAL DATA
G
ENERAL
DATA
T
YPE
DRD–6
2–2, –6, –C, –D
DRD–6
2–3
Measurement range
0 – 60 bar (0 – 870 psig)
(according to the order code)
Overload capability
According to the order code
Accuracy ±0.5%
Medium temperature
(1)
-40 °C … +125 °C (-40 °F … +257 °F)
Ambient temperature
(2)
-25 °C … +85 °C (-13 °F … +185 °F)
(with integrated cable: -5 °C … +70 °C [+23 ° F … +158 °F])
Materials
of the
wetted parts
Sensor
Aluminium oxide ceramic Al
2
O
3
96 %
Sensor sealing
FKM (Viton
®
)
Optional: EPDM, NBR
Process conn.
Stainless steel 1.4404 (316L), Optional: PVDF
Housing
Stainless steel 1.4404 (316L), Optional: Stainless steel 1.4301 (304), PVDF
Output
2-wire 4 – 20 mA
3- wire 0 – 10 V
Power supply (U
Supply
)
(2)
standard version: 8 – 32 V DC,
SIL version: 14 – 28 V DC,
14 – 30 V DC
Load resistance
R
max
=[(U
Supply
-U
Supply min.
)/0.02 A], [
Ω
]
R
min
= 10 k
Ω
Process connection
¾" BSP
Electrical connection
ISO 4400, M12x1 (4-pin), integrated cable
Ingress protection
IP65 (ISO 4400), IP67 (M12x1), IP68 (integrated cable with ventilation tube)
Electric protection
Class III (SELV)
Mass
~0.15 kg (~0.33 lb)
(1)
Permissible medium temperature for PVDF process connection must be between -30 °C … +60 °C (-22 °F … +140 °F)
(2)
For information of Ex certified devices, see Special data for Ex certified models’ chart
2.1
A
CCESSORIES
User’s manual
Warranty Card
EU Declaration of Conformity
S
PECIAL DATA FOR EX CERTIFIED MODELS
(
ONLY FOR
4
–
20
mA
/
2-
WIRE
)
Type
DRD–6
–6 Ex,
DRD–6
–D Ex
Ex marking
(1)
(Pending.)
Ex reference document number
(Pending.)
Ex power supply
10 – 28 V DC
Intrinsically safety data
U
imax
= 28 V DC, I
imax
= 93 mA, P
imax
= 660 mW, C
i
≈
0 nF, L
i
≈
0 µH
The maximum interior capacity between the power supply connections and the metal housing is 27 nF.
Permissible medium temperature,
atmospheric pressure
Zone 0:
-20 °C … +60 °C (-4 °F … +140 °F), 0.8 bar
≤
p
atm
≤
1.1 bar
Zone 1, 2:
-20 °C … +70 °C (-4 °F … +158 °F)
Connecting cable
(in case if the device equipped with an
integrated cable)
Cable capacitance: 160 pF/m
Cable inductivity: 1µH/m
(1)
Only in case of stainless steel process connection!
2.2
O
RDER CODE
(N
OT ALL COMBINATIONS POSSIBLE
!)
NIPRESS D
R
D – 6 2 –
M
EASURING
METHOD
C
ODE
P
ROCESS
CONNECTION
C
ODE
R
ANGE
(1)
/
N
OMINAL
PRESSURE
(bar)
C
ODE
A
CCURACY
C
ODE
O
UTPUT
/
E
X
C
ODE
Gauge
R
¾"
BSP
D
0
– 0.4
/
1
3
0.5%
2
4
– 20 mA, 2-wire
2
0
– 0.6
/
2
4
0
– 10 V, 3-wire
3
0
– 1
/
2
5
4
– 20 mA, 2-wire / Ex ia
(2)
6
0
– 1.6
/
4
6
4
– 20 mA, 2-wire, SIL2
(2)
C
0
– 2.5
/
4
7
4 – 20 mA, 2-wire,
SIL2 / Ex ia
(2)
D
(1)
Custom measuring range, based on prior negotiations
(2)
Ex or SIL versions are available upon special request
(3)
Only available with stainless steel process connection
Note:
In case of non-standard applications, the sensor type, sealing and housing,
as well as the requirement for filling with food compatible oil has to be
specified in the order!
0
– 4
/
10
8
0
– 6
/
20
9
0
– 10
/
20
A
0
– 16
/
40
B
0
– 25
/
40
C
0 – 40
(3)
/ 100
D
0 – 60
(3)
/ 200
E
Dodávate
ľ
:
MICROWELL spol. s r. o.
SNP 2018/42, 927 00 Ša
ľ
a
Tel.: (+421) 31/ 770 7585, 770 7587
E-mail: [email protected]
http://www.microwell.sk