NIVELCO NIPRESS DRD-6 Series User Manual Download Page 1

drd6322a0600h_01 - 1/3  

Thank you for choosing NIVELCO instrument 

We are convinced that you will be satisfied with our product!

 

1. APPLICATION 

NIPRESS D-600

 series pressure transmitters with ceramic flush sensor measure pressure and convert it into 

voltage and current output can be used in either 2- and 3-wire systems. The

 

D-600 transmitters are suitable for 

the pressure measurement of fluids, oils, and gases. Due to their flush sensor, they are ideal for the measurement 
of viscose and polluted media. For using in aggressive media, we recommend PVDF process connection. 
Transmitters can be delivered with plug in display 

UNICONT PLK-501

 enabling on site reading. Ex or SIL 

versions are also available. 

2. TECHNICAL DATA 

G

ENERAL

 

DATA

 

T

YPE

 

DRD–6

2–2, –6, –C, –D 

DRD–6

2–3 

Measurement range 

0 – 60 bar (0 – 870 psig) 

(according to the order code) 

Overload capability 

According to the order code 

Accuracy ±0.5% 
Medium temperature 

(1)

 

-40 °C … +125 °C (-40 °F … +257 °F) 

Ambient temperature 

(2)

 

-25 °C … +85 °C (-13 °F … +185 °F) 

(with integrated cable: -5 °C … +70 °C [+23 ° F … +158 °F]) 

Materials 

of the 

wetted parts 

Sensor 

Aluminium oxide ceramic Al

2

O

3

 96 % 

Sensor sealing 

FKM (Viton

®

Optional: EPDM, NBR 

Process conn. 

Stainless steel 1.4404 (316L), Optional: PVDF 

Housing 

Stainless steel 1.4404 (316L), Optional: Stainless steel 1.4301 (304), PVDF 

Output 

2-wire 4 – 20 mA 

3- wire 0 – 10 V 

Power supply (U

Supply

)

(2)

 

standard version: 8 – 32 V DC, 

SIL version: 14 – 28 V DC, 

14 – 30 V DC 

Load resistance

 

R

max

=[(U

Supply

-U

Supply min.

)/0.02 A], [

R

min

= 10 k

 

Process connection 

¾" BSP 

Electrical connection 

ISO 4400, M12x1 (4-pin), integrated cable 

Ingress protection 

IP65 (ISO 4400), IP67 (M12x1), IP68 (integrated cable with ventilation tube) 

Electric protection 

Class III (SELV) 

Mass 

~0.15 kg (~0.33 lb) 

(1)

 Permissible medium temperature for PVDF process connection must be between -30 °C … +60 °C (-22 °F … +140 °F)  

(2)

 For information of Ex certified devices, see Special data for Ex certified models’ chart

 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
2.1

 

A

CCESSORIES

 

 

User’s manual 

 

Warranty Card 

 

EU Declaration of Conformity

S

PECIAL DATA FOR EX CERTIFIED MODELS 

(

ONLY FOR 

4

 

 

20

 

mA

 

/

 

2-

WIRE

Type 

DRD–6



–6 Ex, 

 DRD–6



–D Ex 

Ex marking

(1)

 

(Pending.) 

Ex reference document number 

(Pending.) 

Ex power supply 

10 – 28 V DC 

Intrinsically safety data 

U

imax

 = 28 V DC, I

imax

 = 93 mA, P

imax

 = 660 mW, C

i

 

 0 nF, L

i

 

 0 µH 

The maximum interior capacity between the power supply connections and the metal housing is 27 nF. 

Permissible medium temperature,  
atmospheric pressure 

Zone 0:  

-20 °C … +60 °C (-4 °F … +140 °F), 0.8 bar 

 p

atm

 

 1.1 bar 

Zone 1, 2: 

-20 °C … +70 °C (-4 °F … +158 °F) 

Connecting cable  
(in case if the device equipped with an 
integrated cable) 

Cable capacitance: 160 pF/m

 

Cable inductivity: 1µH/m 

(1)

 Only in case of stainless steel process connection! 

2.2

 

O

RDER CODE 

(N

OT ALL COMBINATIONS POSSIBLE

!)

 

NIPRESS D 

 

D  – 6    2 –   

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

M

EASURING 

METHOD

 

C

ODE

 

 

P

ROCESS

 

CONNECTION

 

C

ODE

 

  R

ANGE

(1)

 

/

 

N

OMINAL 

PRESSURE 

(bar) 

C

ODE

 

  A

CCURACY

 

C

ODE

 

  O

UTPUT 

/

 

E

X

 

C

ODE

 

Gauge 

 ¾" 

BSP 

 0 

– 0.4 

 0.5% 

 4 

– 20 mA, 2-wire 

 

 

 

 

 

 0 

– 0.6 

   

 

 0 

– 10 V, 3-wire 

 

 

 

 

 

 0 

– 1  

   

 

 4 

– 20 mA, 2-wire / Ex ia

(2)

 

 

 

 

 

 

 0 

– 1.6 

   

 

 4 

– 20 mA, 2-wire, SIL2

(2)

 

 

 

 

 

 

 0 

– 2.5 

   

 

 

4 – 20 mA, 2-wire,

 

SIL2 / Ex ia

(2)

 

(1)

Custom measuring range, based on prior negotiations 

(2)

Ex or SIL versions are available upon special request 

(3)

Only available with stainless steel process connection 

 

Note: 

In case of non-standard applications, the sensor type, sealing and housing, 

as well as the requirement for filling with food compatible oil has to be 

specified in the order! 

 0 

– 4 

10 

   

 

 

 

 

 0 

– 6 

20 

   

 

 

 

   

 0 

– 10 

20 

   

 

 

 

 

 0 

– 16 

40 

   

 

 

 

 

 0 

– 25 

40 

   

 

 

 

 

 

0 – 40

(3)

 / 100 

D  

 

0 – 60

(3)

 / 200 

 

 

 

 

Dodávate

ľ

MICROWELL spol. s r. o. 

SNP 2018/42, 927 00 Ša

ľ

Tel.: (+421) 31/ 770 7585, 770 7587 

E-mail: [email protected]   

http://www.microwell.sk 

Reviews: