RECALLING STORED DATA
1. Press memory M1 button or memory M2 button to see stored data.
The average of the stored result in the selected bank is displayed.
The latest result stored is displayed when there is only a single result and
the clock display remains when there are no results stored.
The memory data is displayed for approximately 30 seconds. Approximately
30 seconds after memory button is released, the apparatus will turn off.
2. Press memory button to move to next data.
The indication at the top of the display alternately changes from memory
data number to date and to time.
The memory data number 1 is the latest among the stored data in the
selected memory bank.
3. Every depression of memory button switches among the memory data.
MEMORY FUNCTION
ERROR DISPLAYS AND TROUBLESHOOTING
As memory button is pressed, the memory data number increases; the
bigger the number, the older the result.
DELETING STORED DATA
1. Show the individual result to be erased or the average of a memory
bank to clear all the memory bank data.
2. Press and hold down either one of memory buttons: M1 button to
erase specified data in memory bank M1, and M2 button to erase
specified data in memory bank M2.
The displayed data starts flashing.
3. Hold down the button until nothing but the memory bank number
is displayed.
OVER-PRESSURIZATION
The pressurization was not enough to make a measurement even
though the cuff was inflated to the maximum value because of
wrist movement or talking during measurement. Do not move
during measurement.
MEASUREMENT ERROR
Measurement could not be made because of moving or talking
during measurement. Remain still and quiet during
measurement.
INFLATION ERROR
The cuff is not properly applied. Properly apply the cuff.
If Err-2 appears after re-trying to apply the cuff properly, assume
the cuff is damaged. Contact your dealer.
DEFLATION ERROR
Movement or talking was detected during measurement. Turn off
the power and make another measurement while being quiet
and still.
-15-
Summary of Contents for WS-820
Page 65: ...WS 820 15 125 215 0 300 mmHg 40 160 2 65 66 67 68 69 69 70 71 72 73 74 WS 820 65...
Page 67: ...20 C 10 mmHg 67...
Page 69: ...1 2 5 10 mm START START 1 2 3 Fig 3 4 3 Fig 4 5 300 69...
Page 70: ...1 START O 2 M1 M2 3 START 3 START 70...
Page 71: ...M1 M2 30 30 1 1 2 40 mmHg START 50 mmHg 190 mmHg 300 mmHg Fig 6 8 5 71...
Page 72: ...1 2 M1 1 M2 M2 3 2 30 30 2 1 3 72...
Page 73: ...1 2 START START 5 73...
Page 75: ......