GENERAL PRODUCT INFORMATION
Accuracy
: ±3 mmHg (cuff air pressure)
±5% of reading (pulse rate)
Inflation
: Automatic inflation
Deflation
: Electric control valve
Exhaust
: Automatic exhaust valve
Power Supply
: 7.0 VDC AC adaptor (MODEL UM318-0725) or 4.8 VDC rechargeable
NiMH battery
Memory
: seven measurement results
Operating Environment
: +10°C to+40°C, 15% to 85% RH (noncondensing)
Storage Environment
: -20°C to+60°C, 15% to 85% RH (noncondensing)
Coverage Arm Circumference
Regular Size Cuff
: 22 to 32 cm
Large Size Cuff
: 32 to 42 cm
Main Unit
Weight
: Approx. 1350 g, without AC adaptor or battery
Size
: 136 x 228 x 276 mm (W x D x H)
Protection against electric shock
: Internally powered equipment / Class II equipment, Type BF applied part
Classification of safety level when used in air/flammable anesthetic gas, oxygen, nitrous oxide/
flammable anesthetic gas.
: Not suitable for use in these environments
Mode of operation
: Continuous operation
Classification
: Class II / Internally powered equipment
Expected service life
: 5 years
Key to Symbols
: Type BF equipment
: Class II equipment
: Important; read operating instructions.
: The used electrical and electronic products are not household waste.
Follow your national/local recycling rules to dispose of them properly. In
the EU countries, please refer to waste management symbol(s) marked
on the package or the instrument.
*Accuracy is guaranteed with the measured values that are within the measuring range.
• This device complies with EN1060-1:1995+A2:2009 Non-invasive sphygmomanometers Part 1:
General requirements and EN1060-3:1997+A2:2009 Non-invasive sphygmomanometers Part 3:
Supplementary requirements for electro-mechanical blood pressure measuring system
• For the measurement in AUTO mode, the measurement accuracy of the device has been proven
according to ISO 81060-2 protocol. In the clinical study, K5 was used for the determination of diastolic
pressure values at all auscultatory measurements.
• This device is intended for use in the environment with one atmospheric pressure.
Specifications are subject to change without notice due to improvements in performance.