Nissei DM-3000 Instructions Manual Download Page 5

- 5 -

REGULAR SIZE CUFF

for arm circumference of 22 to 32 cm

AC ADAPTOR

INSTRUCTION

MANUAL

RECHARGEABLE

NICKEL METAL HYBRID

(NiMH) BATTERY

LARGE SIZE CUFF

for arm circumference of 32 to 42 cm

AIR TUBE

AIR PLUG

ADAPTOR PLUG

BATTERY PLUG

AIR CONNECTOR

AIR CONNECTOR

Order No.: NPDM3000-010

CUFF SHELL : NPDM3000-030

BLADDER : NPDM3000-040

AIR PLUG : B112784-1

Order No.: NPDM3000-011

CUFF SHELL : NPDM3000-031

BLADDER : NPDM3000-041

AIR PLUG : B112784-1

Order No.:  NPDM3000-001(EU)
 NPDM3000-002(UK)

Order No.:  NPDM3000-005

Model 

: DM-3000

Operating Principle 

:  Oscillometric method (AUTO measurement) / auscultation (MANUAL 

measurement)

Indicator 

:  300 digits column LCD and 10 digits segment LCD

Pressure Indicating Range  :  0 to 300 mmHg (cuff pressure)
Measuring Range 

:  50 to 250 mmHg (systolic) 

40 to 180 mmHg (diastolic) 

40 to 160 bpm (pulse rate)

DM-3000 is an automatic sphygmomanometer intended for measurement of systolic and diastolic 

blood pressure and pulse rate in adult patients, i.e., age 12 and above. This product is not designed 

for neonatal use and inaccurate readings may result if it is used on children.

PRODUCT FEATURES

COLUMN LCD and SEGMENT LCD

The pressure can be read in the column LCD height and as digital value in the segment LCD. 

"AUTO" and "MANUAL" Modes of Measurement

Determination of blood pressure can be made either automatically by the oscillometric method or 

manually by the Auscultatory method using a stethoscope.

Regular and Large Size Cuffs

The regular size cuff will accommodate an upper arm circumference range of approximately 22 to 32 

centimeters and the large size cuff will accommodate that of approximately 32 to 42 centimeters.

Operation with AC Adaptor or Rechargeable Battery

DM-3000 is powered by either an AC adaptor or rechargeable nickel metal hybrid (NiMH) battery.

PRODUCT SPECIFICATIONS

COMPARISON TABLE

MANUAL

MODE

cuff inflation

deflation rate

saving result

AUTO

maintained at 4.0 to 

4.9 mmHg/sec
maintained at preset 

rate: 2.5, 4.5 or 6.5 

mmHg/sec

automatically performed to 

preset pressure value
automatically performed to 

preset pressure value or 

manually performed with 

inflation bulb*

seven measured results 

automatically saved in memory.

 * Inflation bulb is not included in the product.

GENERAL PRODUCT INFORMATION

Summary of Contents for DM-3000

Page 1: ...purposes other than those described herein It is essential to read and understand the entire manual before use CONTENTS PARTS AND PRODUCT COMPONENTS 3 GENERAL PRODUCT INFORMATION 5 WARNING PRECAUTION...

Page 2: ...n the cuff pressure is gradually changed The blood pressure is determined according to the pressure values at appearance and disappearance of the Korotkov sounds Intended purpose and Indication for us...

Page 3: ...ARK DEFLATION MARK PULSE MARK BATTERY MARK POWER MARK PRESET PRESSURE CONTROL AC INDICATOR MODE SELECTOR SET MEMORY BUTTON START STOP BUTTON MAIN POWER SWITCH OFF ON AC ADAPTOR JACK BATTERY COMPARTMEN...

Page 4: ...cumference of 32 to 42 cm AIR TUBE AIR PLUG ADAPTOR PLUG BATTERY PLUG AIR CONNECTOR AIR CONNECTOR Order No NPDM3000 010 CUFF SHELL NPDM3000 030 BLADDER NPDM3000 040 AIR PLUG B112784 1 Order No NPDM300...

Page 5: ...ion of blood pressure can be made either automatically by the oscillometric method or manually by the Auscultatory method using a stethoscope Regular and Large Size Cuffs The regular size cuff will ac...

Page 6: ...ternally powered equipment Expected service life 5 years Key to Symbols Type BF equipment Class II equipment Important read operating instructions The used electrical and electronic products are not h...

Page 7: ...to interfered blood flow from cuff inflation make sure that AIR TUBE is not kinking or compressed before measurement Otherwise cuff inflation may not be conducted properly and prolonged Make sure that...

Page 8: ...rwise injury may be resulted Do not make measurements repeatedly to avoid harmful injury due to interfered blood flow from cuff inflation Do not drop or strike the product Make sure not to expose it t...

Page 9: ...e screw Connect BATTERY PLUG to BATTERY CONNECTOR inside the battery compartment and insert the battery Put back BATTERY COMPARTMENT COVER and close it by inserting and tightening the screw Connect AC...

Page 10: ...o be charged Avoid charging the battery if it is not completely exhausted Do not forcibly remove the battery or pull out the plugs holding cords and cables Use manufacturers dedicated AC adaptor and b...

Page 11: ...ze of cuff for your patient Find a chair and a table so that the patient can be comfortably seated with patient s back and arm supported Legs should not be crossed and feet kept flat on the floor The...

Page 12: ...is rapidly exhausted from the cuff when measurement is completed Blood pressure and pulse rate are displayed Turn off the device by pressing side of MAIN POWER SWITCH The device will go into energy sa...

Page 13: ...the segment display The deflation rate can be selected from 2 5 4 5 and 6 5 mmHg sec The set deflation rate returns to the default value of 4 5 when the device is turned off Apply the cuff and stetho...

Page 14: ...e will go into energy saver mode after approximately 3 minutes even if the SWITCH has not been turned off See page 10 Manually pressurizing the cuff with inflation bulb Disconnect AIR PLUG from DM 300...

Page 15: ...ed first 2 Press M button to recall stored data one by one Figure following No indicates sequence of data the larger the number the older the data However maximum data can that be stored is 7 Deleting...

Page 16: ...n off the device or press START STOP BUTTON or SET MEMORY BUTTON to activate the device from sleeping mode OVER PRESSURIZATION The cuff was inflated to the maximum pressure because of movement of body...

Page 17: ...R Battery is exhausted ENERGY SAVER MODE The device was left turned on and has not been operated for over approximately 3 minutes When the preset pressure value was not high enough to make measurement...

Page 18: ...not operate when PRESET PRESSURE is set to PUMP OFF Turn PRESET PRESSURE CONTROL and select appropriate pressure between 100 and 280 mmHg for automatic inflation AC INDICATOR also indicates whether t...

Page 19: ...g Drinking Alcohol Smoking Measurement posture Recent Urination or Bowel Movement Room temperature Changes in the surroundings such as movement or noise etc Blood pressure measured at hospital or clin...

Page 20: ...r There should be no speaking or moving during blood pressure measurement Otherwise correct measurement results cannot be obtained The ambient temperature should be approximately 20 C for blood pressu...

Page 21: ...t is stored in the environment with ambient temperature above 40 C or below 10 C please leave it for at least 2 hours before taking a measurement Remove the battery if the device is to be stored for a...

Page 22: ...d holding START STOP BUTTON down Release the button after the initial display all segment display disappears Two s are displayed and DM 3000 is in CHECKING MODE Inflate the cuff with an inflation bulb...

Page 23: ...equipment and MRI magnetic resonance imaging equipment etc Use of the device adjacent to or stacked with other equipment must be avoided because it could result in improper operation Use of accessorie...

Page 24: ...damage caused by natural disaster violent action or war Also cuffs including cuff shells bladders tubes and plugs are not covered by this warranty except for defects in manufacture Purchaser shall be...

Reviews: