If cuff inflation was not sufficient, press and hold down
START/STOP BUTTON once deflation has started and the cuff is
pressurized again. Pressurization will continue while the button
is pressed. Pressure does not exceed 290 mmHg.
Press START/STOP BUTTON when blood pressure is taken to
exhaust residual air rapidly from the cuff.
Turn off the device by pressing " " side of MAIN POWER SWITCH.
The device will go into energy saver mode after approximately 3
minutes even if the SWITCH has not been turned off. See page 10.
Manually pressurizing the cuff with inflation bulb
Disconnect AIR PLUG from DM-3000 original cuff and connect it to the cuff to
use. Insert AIR PLUG into AIR CONNECTOR. Push the MODE SELECTOR
upward and set to "MANUAL". Set the PRESET PRESSURE CONTROL to
"PUMP OFF" and press START/STOP BUTTON. Start measurement after
INFLATION MARK is displayed. To use control valve of the inflation bulb
most effectively, set the deflation rate at 2.5 mmHg on DM-3000.
* Inflation bulb not included with DM-3000.